Troubleshooting; Fault Messages - Nonin RespSense LS1R-9R Operator's Manual

Capnography monitor
Table of Contents

Advertisement

Troubleshooting

Fault Messages

RespSense has built-in self-diagnostics for detection of fault conditions. Detected fault
conditions are presented as messages on the touch panel display. The fault conditions are either
operator- or system-generated. The table below lists common messages, descriptions, and
advice on actions to take.
If the problem persists, contact Nonin Technical Service.
Message
OCCLUSION
Sample line or cannula occlusion.
Incorrect placement of the
moisture trap.
Clogged filter.
Sample line is not properly applied
to the patient.
Sample line or Nafion tubing is not
connected to the moisture trap.
BATT LOW
Battery is low. Monitor will run for
approximately 20 minutes.
DISP ERROR
The display is not showing any
parameters.
Description
Action
Remove obstruction. Replace the
sample line or cannula.
Reposition the moisture trap.
Replace the filter.
Verify sample line placement.
Verify connection to moisture trap.
Plug the power supply into a power
outlet and charge the monitor.
If the monitor continues to show
BATT LOW message after recharging,
contact Nonin Technical Service as the
battery may need replacement. The
battery is integral to the device and
cannot be replaced by the operator.
Turn off the monitor and then turn on
again. If the problem persists, contact
Nonin Technical Service.
27