Table 9. Electromagnetic Immunity - Nonin RespSense LS1R-9R Operator's Manual

Capnography monitor
Table of Contents

Advertisement

Immunity Test
This device is intended for use in the electromagnetic environment specified below. The user of this device
Electrostatic
±6 kV contact
Discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical Fast
± 2 kV for power
Transient/Burst
supply lines
IEC 61000-4-4
± 1 kV for input/
output lines
Surge
± 1 kV differential
mode
IEC 61000-4-5
± 2 kV for common
mode
Voltage dips, short
± 5% U
interruptions, and
in U
voltage variations on
± 40% U
power supply input
in U
lines
± 70% U
IEC 61000-4-11
in U
± 5% U
in U
Power Frequency (50/
3 A/m
60 Hz) Magnetic Field
IEC 61000-4-8
Note: U
is the AC mains voltage before application of the test level.
T
Table 9: Electromagnetic Immunity
IEC 60601 Test
Compliance Level
Level
should ensure that it is used in such an environment.
±6 kV contact
±8 kV air
± 2 kV for power
supply lines
± 500V for input/
output lines
± 1 kV differential
mode
± 2 kV for common
mode
(>95% dip
± 5% U
T
) for 0.5 cycle
U
) for 0.5 cycle
T
T
(60% dip
± 40% U
T
) for 5 cycles
U
) for 5 cycles
T
T
(30% dip
± 70% U
T
) for 25 cycles
U
) for 25 cycles
T
T
(>95% dip
± 5% U
T
) for 5 cycles
U
) for 5 cycles
T
T
3 A/m
Environment—Guidance
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
(>95% dip in
Mains power quality should be that of
T
a typical commercial or hospital
environment. If the user of the device
(60% dip in
T
requires continued operation during
power mains interruptions, it is
(30% dip in
recommended that the device be
T
powered from an uninterruptible
(>95% dip in
power supply or battery pack.
T
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
Electromagnetic
35