Page 1
Operating Manual Laboratory Centrifuge with Heating System 6-16HS Spincontrol S Translation of the original operating manual From serial no. 145877 sb-kue Version 06/2012 Rev. 1.0 6-16HS; ;Spincontrol S Please retain for later use!
Page 2
6-16HS Version 06/2012, Rev. 1.0 • sb-kue Translation of the original operating manual...
6-16HS Table of contents General information ......................8 Importance of the operating manual .................. 8 Intended use ........................8 Warranty and liability ......................8 Copyright .......................... 9 Standards and regulations ....................9 Scope of Supply ........................ 9 Layout and mode of operation ..................10 Layout of the centrifuge ....................
Page 6
6-16HS Table of contents Storage and transport ....................... 23 Dimensions and weight ....................23 Storage conditions ......................23 Notes on transport ......................23 Packaging ........................24 Transport safety device ....................24 Set-up and connection ...................... 25 Installation site ........................ 25 Power supply ........................
Page 7
6-16HS Table of contents Malfunctions and error correction ..................49 General malfunctions ...................... 49 7.1.1 Emergency lid release ....................50 Table of error codes ......................51 Service contact ....................... 52 Maintenance and service....................53 Maintenance ........................53 8.1.1 Centrifuge ........................53 8.1.2...
They are solely intended for this purpose. Any other use beyond this area of application is regarded as improper use. Sigma Laborzentrifugen GmbH cannot be held liable for any damage resulting from such improper use.
Pos: 17 / 100 Sigma /110 BA Ze ntrifu gen (Proj ekte )/6- 16S_ 6-1 6HS/020 Aufba u u nd Wirk ungsw eise/ 020 -00 10- 001 0 F unktio ns- und Be dien elem ente 6- 16S @ 12\ mod _134 034 852 824 5_6 8.doc...
Pos: 24 / 100 Sigma /100 BA Ze ntrifu gen (Sta nda rdm odul e)/0 20 Auf bau und Wirku ngsweis e/02 0-0 020 -00 20 Anwen dun gsbe reich @ 7\ mod _13 068 365 440 60_ 68.d...
Pos: 34 / 100 Sigma /100 BA Ze ntrifu gen (Sta nda rdm odul e)/0 30 Sich erh eit/0 30- 0020 Sym bol- und Hi nweise rklär ung en @ 7\ mod _130 630 280 997 8_6 8.doc...
Pos: 80 / 100 Sigma /100 BA Ze ntrifu gen (Sta nda rdm odul e)/0 40 Lage run g u nd Tra nspo rt/0 40- 002 0 L age rbe ding ung en @ 7\m od_ 130 630 8899 781 _68 .docx...
6-16HS 6 Using the centrifuge Once a day or after 20 cycles, the rotor tie-down screw must be loosened by some turns, and the rotor must be lifted and fastened again. This ensures a proper connection between the rotor and the motor shaft.
6-16HS 6 Using the centrifuge Pos: 112 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 060 Betrie b_Spinc ont rol L +S/06 0-0 020 -00 20- 004 0 Ad apte r - nur L+S @ 7\ mod _130 640 367 744 1_6 8.doc x @ 4 359 7 @ 44 @ 1 6.2.2.4...
6-16HS 6 Using the centrifuge • When using smaller bag systems or in case of incompletely filled blood bags adapters (e.g. part no. 17750) have to be inserted together with the blood bag systems. This will avoid slipping of the bags which could result in an imbalance.
6-16HS 6 Using the centrifuge Pos: 123 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 060 Betrie b_Spinc ont rol L +S/Spinco ntr ol S ( Zent 7)/ 060 -00 30- 0020 M anu eller B etrie b @ 1 0\m od_ 132 869 710 5069 _68 .docx @ 5 802 8 @ 3 @ 1 6.3.2...
The standard menu is displayed a few seconds after the centrifuge has been switched on. The following parameters can be displayed and modified. As a default setting of the 6-16HS, the petroleum program "ASTM oil" is stored in the system (see the "program selection list" in this chapter). NOTE Fig.
Page 34
6-16HS 6 Using the centrifuge Runtime The set runtime is displayed in the upper section of this field, with the remaining runtime shown below. The runtime is defined as the period from the start of the centrifuge to the beginning of the deceleration phase. The maximum value is 99 h 59 min 59 sec.
Page 35
6-16HS 6 Using the centrifuge Rotor: rotor selection list/ automatic rotor identification This field shows the rotor that is currently being used. • Select the field "Rotor" and confirm the selection. A list with all of the possible rotors without buckets is displayed.
Page 36
6-16HS 6 Using the centrifuge Progr.: program selection list This field in the standard menu shows the program that is currently loaded. When the field is selected, the program list is displayed (for information on how to work with the programs, chapter 6.3.3 - "Program mode").
Page 37
6-16HS 6 Using the centrifuge corresponding RCF value. The deceleration (brake) and acceleration curves correspond to curve 9 and the runtime field indicates “infinite” (continuous run). Fig. 15: "RAPID_TEMP" program The program will only be loaded if the actual temperature is below the set temperature.
6-16HS 6 Using the centrifuge The delta T temperature monitoring system (see chapter (see chapter 6.3.2.6 - "Parameters menu"), Process) remains inactive as long as the "RAPID_TEMP" program is active. NOTE If the "RAPID_TEMP" program is used, the temperature of the unloaded aluminium bucket will be displayed.
Page 39
6-16HS 6 Using the centrifuge If the standstill heating is activated, the centrifuge starts to preheat after it is switched on. In the runtime field, the symbol " " is displayed. The lid must be closed. The centrifuge is equipped with the "RAPID_TEMP" program. This program is used to precool the rotor chamber quickly under defined conditions (see chapter 6.3.2.5 - "Standard menu", Program list).
Page 40
6-16HS 6 Using the centrifuge Temperature monitoring The control system includes a temperature monitoring function. If the set temperature difference with regard to the set value deviates from the actual temperature difference, this function will stop the centrifuge and issue an error message.
6-16HS 6 Using the centrifuge If changes are made to a blocked function, the system will ask for the code prior to executing the change. Unblocking a function • Select the button "Deactivate code". • Enter the code and confirm the entry.
Page 42
6-16HS 6 Using the centrifuge Program rotation See chapter 6.3.3.4 - "Automatic program rotation" Runtime as of set speed If this function is active, the runtime will not be measured until the set speed is reached. The clock symbol " !" will be displayed in the runtime field.
Page 43
6-16HS 6 Using the centrifuge Invert If this function is activated, the display switches from the standard setting with a bright background and dark writing to a dark background with bright writing. System Cycles This field shows the number of cycles as well as the runtime of the rotor and buckets that are currently being used.
6-16HS 6 Using the centrifuge Pos: 138 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 060 Betrie b_Spinc ont rol L +S/Spinco ntr ol S ( Zent 7)/ 060 -00 30- 0020 -0 090 Me nü Hilf e @ 1 0\m od_ 132 8697 107 366 _68 .docx @ 58 100 @ 4 4 @ 1 6.3.2.8...
6-16HS 6 Using the centrifuge Pos: 142 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 060 Betrie b_Spinc ont rol L +S/Spinco ntr ol S ( Zent 7)/ 060 -00 30- 0030 Pro gra mm bet rieb @ 10\ mo d_1 3286 971 080 70_ 68.d ocx @ 581 24 @ 44 @ 1 6.3.3...
6-16HS 6 Using the centrifuge • When the program name is complete, select the option "OK" and confirm it. The program is now saved. Fig. 22: Entering the program name before saving Pos: 147 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1...
6-16HS 6 Using the centrifuge Pos: 150 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 060 Betrie b_Spinc ont rol L +S/Spinco ntr ol S ( Zent 7)/ 060 -00 30- 0030 -0 040 Auto matisc he Pro gr amm rot ation @ 1 2\mo d_1 336 718 285 484 _68. docx @ 64 401 @ 44 @ 1 6.3.3.4...
6-16HS 6 Using the centrifuge Pos: 152 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 060 Betrie b_Spinc ont rol L +S/Spinco ntr ol S ( Zent 7)/ 060 -00 30- 0030 -0 050 Opti one n zur Ein- und Aus gab e vo n Dat en @ 10\ mod _13 286 971 0946 1_6 8.d ocx @ 5816 4 @ 22 @ 1 6.3.3.5...
6-16HS 8 Maintenance and service Pos: 172 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 080 Wart ung und Inst andh altu ng/0 80- 001 0-0 020 Z ube hör @ 7\ mod _13 067 392 3103 7_6 8.d ocx @ 438 27 @ 444 4 @ 1 8.1.2...
6-16HS 8 Maintenance and service Pos: 178 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 080 Wart ung und Inst andh altu ng/0 80- 001 0-0 030 Rot or, Bec her un d Vielfac hträ ger (a uße r Basic ) @ 7 \mo d_1 306 741 3025 33_ 68. docx @ 43 848 @ 33 33 @ 1 8.1.3...
6-16HS 8 Maintenance and service In order to completely remove the glass particles and metal dust from the rotor chamber: • Grease the upper third of the rotor chamber with e.g. Vaseline. • Then, let the rotor rotate for a few minutes at a moderate speed (approx.
6-16HS 8 Maintenance and service Pos: 192 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 080 Wart ung und Inst andh altu ng/0 80- 004 0 Rüc ksen dun g de fekte r T eile @ 7\m od_ 130 674 407 898 8_68 .docx @ 4 390 4 @ 2 22 @ 1...
Pos: 196 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 090 Entso rgun g/0 90- 001 0 Ents org ung de r Z entrif uge -- --- -- --- --- -- --- -- --- -- --- --- -- --- -- --- --- -- @ 7\ mod _13 067 454 030 22_6 8.d ocx @ 439 18 @ 22 @ 1 Disposal of the centrifuge • In accordance with the directive 2002/96/EC, SIGMA centrifuges are marked with the symbol shown to the left. This symbol means that it is not permissible to dispose of the unit among household waste.
(e.g. circuit diagrams) and the safety data sheets of the manufacturers of refrigerant and lubricants are not attached to this documentation. You can order these documents at www.sigma-zentrifugen.de by way of our "service request" form. Pos: 206 /01 0 Univ ersal mod ule/ Ab schnitt swechs el @ 0 \mo d_1 202 124 514 062_ 0.d ocx @ 418 @ 1 @ 1...
6-16HS 11 Appendix Pos: 208 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 110 Anha ng/1 10 A nhan g= == === == == === == == === == == === == == == === = @ 7\m od_ 130 6748 620 237 _68. docx @ 43 955 @ 1 @ 1 11 Appendix Pos: 209 /10 0 Sigm a/11 0 BA Z entrif uge n (Pr ojekt e)/6 -16S _6- 16HS/11 0 Anh ang/ 110 -00 10 Zub ehö...
Page 64
6-16HS 11 Appendix Part no. Description Max. speed Max. (rpm) gravitational field (x g) 12170 Angle rotor for 12 x 15 ml culture tubes e.g. no. 15115, incl. hermetic aluminium lid no. 17859 Radius max. 10.0 cm, radius min. 4.6 cm, angle 25°...
Page 65
6-16HS 11 Appendix Part no. Description Max. speed Max. (rpm) gravitational field (x g) 13233 Multiple carrier for 50 tubes approx. 5 ml, max. Ø 12.5 x 70 - 90 mm, e.g. RIA tube no. 15060, 1 set = 2 pcs.
Page 66
6-16HS 11 Appendix Adapters and stainless steel tubes Part no. Description 13000 Adapter for reaction vials 0.25/0.4 ml no. 15014, suitable for 11148, 12130, 12201, 17353, 17653, 18002, 18003, 1 Set = 2 Stück 13002 Adapter for reaction vials 0.5/0.75 ml, Ø 7.9/10 x 28/31 mm, e.g. no. 15005, suitable for 11148, 12130, 12201, 17353, 17653, 18002, 18003, 1 set = 2 pcs.
6-16HS 11 Appendix Pos: 217 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 110 Anha ng/1 10- 003 0 Besc hleu nigun gs- und B rem skurv en @ 7\ mod_ 130 683 683 832 4_6 8.docx @ 4 404 6 @ 2 22 @ 1 11.3...
Page 70
6-16HS 11 Appendix Linear Curves 98 7 6 5 4 1000 curve 0 curve 1 curve 2 curve 3 curve 4 curve 5 curve 6 curve 7 curve 8 curve 9 Time [sec] Fig. 32: Diagram of linear curves Quadratic curves The deceleration curves are inverted images of the acceleration curves and are assigned the same numbers.
6-16HS 11 Appendix Pos: 219 /10 0 Sigm a/10 0 BA Z entrif uge n (St anda rd mod ule)/ 110 Anha ng/1 10- 004 0 Tab elle: Ro tor en u nd Zub ehö r mit abw eichen der Le bens dau er @ 7\m od_ 130 734 285 644 8_68 .docx @ 4 406 1 @ 2 222 @ 1 11.4...
6-16HS 12 Index 12 Index Centrifugation with tubes in compliance with ASTM standard D91/D96 ....... 30 Acceleration ..........38 Centrifuge cannot be started ...... 49 Acceleration curve ....... 38, 69 Centrifuge decelerates during operation ..49 Accessories, cleaning and care ....54 Centrifuge is inclined ........
Page 80
6-16HS 12 Index Declaration of conformity ....... 9, 78 Declaration of decontamination ....59 General conditions ........8 Declaration of Decontamination ....60 Glass breakage .......... 55 Decontamination agent ......54 Glass particles ........... 55 Decontamination agents ......56 Gravitational field ........61 Deformation of tubes ........
Page 81
6-16HS 12 Index Leaks ........... 29, 57 Optical signal ..........43 Lid cannot be opened ........ 49 Option for the input and output of data ..43 Lid error ............. 51 Options for data input and output ....48 Overseas shipping ........23 Lid lock device ...........
Page 82
6-16HS 12 Index Return of centrifuges, spare parts, or Slushing oil ..........54 accessories ..........59 Slushing oil ..........55 Return of defective parts ......59 Solvents ......... 19, 53, 54 Risk of electrical shock....... 17 Sound signal ..........21 Rotor chamber ...........
Page 83
6-16HS 12 Index Temperature monitoring ......40 Useful volume Temperature monitoring system ....21 - volume that is stated for the tube ..29, 67 Temperature unit........42 User interface ..........31 Temperature value not reached ....49 UV radiation ......... 25, 53 Thermal stress ...........