Connecting To Terminals; Connecting To A Plug - HBM HBK U93 Mounting Instructions

Table of Contents

Advertisement

Available languages

Available languages

* TEDS chip only if ordered
Fig. 7.1
U93 pin assignment (six-wire configuration)
Tip
Swap the white and red cable wires over, if you need a negative output voltage at the ampli­
fier under compressive loading.
7.2.1

Connecting to terminals

1. Cut the cable sheath as shown in Fig. 7.2, to gain access to the shield.
2. Connect the shield extensively to the housing ground.
7.2.2

Connecting to a plug

Connect the cable shield extensively to the connector housing.
10
Fig. 7.2
Cable sheath with incision
U93
MECHANICAL INSTALLATION
wh (white)
bk (black)
rd (red)
bu (blue)
gn (green)
gy (gray)
ye (yellow)
100
10
Measurement signal (+) U
Excitation voltage (-) U
Measurement signal (-) U
Excitation voltage (+) U
Sense lead (+)
Sense lead (-) and TEDS*
Cable shield, connected to the housing
2
Cable sheath cut open around the
circumference and pushed
forward by about 2 mm
A
and TEDS*
B
A
B