B-4 Error Messages
Error Messages and Causes
Code
Message
2110
Config error 10, Insufficient
system memory
2111
Config error 11, Invalid
instrument address
2113
Config error 13, Logical
address or IACK switch set
wrong
2129
Config warning, Sysfail
detected
2130
Config error 30, Pseudo
instrument logical address
unavailable
2131
Config error 32, File system
start up failed
2145
Config warning, Non-volatile
RAM contents lost
2148
Config warning, Driver RAM
contents lost
2202
Unexpected interrupt from
non-message based card
2809
Interrupt line has not been
set up
Cause
Too many instruments installed for the amount
of RAM installed in the mainframe. Cannot
configure instruments. Only the system
instrument is started.
A device's logical address is not a multiple of 8
and the device is not part of a combined
instrument.
Duplicate logical addresses set or interrupt
bypass switches set improperly. Only the system
instrument is started.
A device was asserting SYSFAIL on the
backplane during startup.
A physical device has the same logical address as
IBASIC (240)
Insufficient system resources to allow the
IBASIC file system to start.
NVRAM was corrupted or a cold boot was
executed.
Driver RAM was corrupted or a cold boot was
executed.
A register based card interrupted when an
interrupt service routine had not been set up.
A DIAG:INT:ACT or DIAG:INT:RESP
command was executed before setting the
interrupt with DIAG:INT:SET.