STATus :PRESet
Example
:PRESet
Example
:QUESTionable
7-34 System Instrument Command Reference
Reading the Event Register
STAT:OPER:EVEN?
enter statement
STATus:PRESet sets each bit in the enable register (standard operation status
group) to '0'.
Presetting the Enable Register
STAT:PRES
The STATus:QUEStionable commands are supported by the system instrument,
however, they are not used by the System Instrument. Queries of the
Questionable Data condition and event registers will always return + 0.
query if bit(s) is set
preset enable register