:SELect
VXI:SELect < logical_addr> specifies the logical address which is to be used by
many subsequent commands in the VXI subsystem.
Parameters
Parameter
logical_addr
Comments
Example
Select a logical address
:SELect?
VXI:SELect? returnsthe logical address which will be used by many subsequent
commands in the VXI subsystem. If no logical address has been selected, this
query will return -1.
Parameter
Name
Type
numeric
The *RST default value for logical_addr is that no logical address is
selected (i.e., -1). All other commands which require a logical address to
be selected will respond with Error -221 ("settings conflict") if no logical
address is selected.
When a command encounters an Error -240 ("Hardware Error") the
equivalent of a *RST is executed. This will cause the selected logical
address to be set to -1.
Related Commands: VXI:CONF:LADD?
VXI:SEL 64
System Instrument Command Reference 7-63
VXI :SELect
Range of
Values
0 through 255
sets the logical address to be
used by subsequent VXI
subsystem commands to 64.
Default
Units
none