Agilent Technologies E1300B User Manual page 178

75000 series b mainframes
Hide thumbs Also See for E1300B:
Table of Contents

Advertisement

SYSTem:TIME
Parameters
Comments
Example
:TIME?
Example
:VERSion?
Comments
Example
7-50 System Instrument Command Reference
:TIME
SYSTem:TIME < hour> ,< minute> ,< second> sets the E1300/E1301
mainframe's internal clock.
Parameter
Name
hour
minute
second
Related Commands: SYST:DATE, SYST:DATE?, SYST:TIME?
*RST Condition: *RST does not change the Command Module's real
time clock.
Setting the system time
SYST:TIME 14,30,20
SYSTem:TIME? [MAX| MIN,MAX| MIN,MAX| MIN] returns:
When no parameter is sent; the current system time in the form
+ HH,+ MM,+ SS, where HH can be 0 through 23 hours, MM can be 0
through 59 minutes, and SS can be 0 through 60 seconds.
When parameters are sent; the minimum or maximum allowable values
for each of the three parameters. The parameter count must be three.
Querying the system time
SYST:TIME?
input values of hour,min,sec
SYSTem:VERSion? Returns the SCPI version for which this instrument
complies.
The returned information is in the format: YYYY.R; where YYYY is the
year, and R is the revision number within that year.
Related Commands: *IDN?
Determine compliance version for this instrument.
SYST:VERS?
enter statement
Parameter
Type
numeric
must round to 0 to 23
numeric
must round to 0 to 59
numeric
must round to 0 to 60
Range of
Default
Values
Units
set 2:30:20 PM
ask for current time
read back time
Statement enters 1990.0
none
none
none