SYSTem:COMMunicate :SERial[n] [:RECeive] :PACE :THReshold :STARt
:COMMunicate
:SERial[n] [:RECeive]
:PACE :THReshold
:STARt
Parameters
Comments
Example
:COMMunicate
:SERial[n] [:RECeive]
:PACE :THReshold
:STARt?
Comments
Example
7-42 System Instrument Command Reference
SYSTem:COMMunicate:SERial[n][:RECeive]:PACE:THReshold:STARt
< char_count> configures the input buffer level at which the specified interface
may send the XON character (ASCII 11
the RTS line.
Parameter
Parameter
Name
Type
char_count
numeric
To determine the size of the input buffer of the serial interface you are
using, send SYST:COMM:SER[n]:PACE:THR:START? MAX. The
returned value will be the buffer size less one.
STARt must be set to less than
The
THR:STAR command has no effect unless
PACE:PROT XON,
been sent.
Related Commands:
CONT:RTS
*RST Condition: No change
Set interface to send XON when input buffer contains 10 characters.
SYST:COMM:SER0:PACE:PROT XON
SYST:COMM:SER0:PACE:THR:STAR 10
SYSTem:COMMunicate:SERial[n][:RECeive]:PACE:THReshold:STARt?
[MIN | MAX] returns:
The current start threshold if no parameter is sent.
The maximum allowable setting if MAX is sent.
The minimum allowable setting if MIN is sent.
To determine the size of the input buffer of the serial interface you are
using, send SYST:COMM:SER[n]:PACE:THR:START? MAX. The
returned value will be the buffer size.
Return current start threshold
SYST:COMM:SER0:PACE:THR:STAR? query for threshold value
enter statement
), assert the DTR line, and/or assert
16
Range of
Values
1 through 99 for built-in
1 through 8191 for E1324A
STOP.
CONT:DTR IBF, or
PACE:PROT XON | NONE,
statement enters a numeric
value
Default
Units
none
CONT:DTR IBF has
CONT:DTR,