Agilent Technologies E1300B User Manual page 135

75000 series b mainframes
Hide thumbs Also See for E1300B:
Table of Contents

Advertisement

:BOOT[:WARM]
Comments
Example
:COMMunicate
:SERial[0][:OWNer]
Parameters
Comments
Example
DIAGnostic:BOOT[:WARM] causes the System Instrument to restart (re-boot)
using the current configuration stored in non-volatile memory. The effect is the
same as cycling power.
The System Instrument goes through its power-up self tests.
The non-volatile system state is used for configuration wherever
applicable.
Related Commands: DIAG:BOOT:COLD
Booting the System Instrument (warm)
DIAG:BOOT:WARM
DIAGnostic:COMMunicate:SERial[0][:OWNer] < owner > Allocates the
built-in serial interface to the System Instrument, the optional IBASIC
interpreter, or to neither.
Parameter
Parameter
Name
Type
owner
discrete
While the serial interface is allocated to the Command Module
(SYSTem), it can function as the mainframe user interface when
connected to a terminal or computer running terminal emulation
software.
When the built-in serial interface is allocated to IBASIC, it is controlled
only by IBASIC. The serial interface is given a select code of 9, and any
RS-232 device connected to the (Command Module) RS-232 port is
programmed accordingly.
If the built-in serial interface is not needed, specifying NONE will release
memory for use by other instruments.
Once the new serial interface owner has been specified
(DIAG:COMM:SER:OWN), the change will not take effect until you
re-boot (warm) the system.
Related Commands: DIAGnostic:COMMunicate:SERial[:OWNer]
Give the serial interface to IBASIC.
DIAG:COMM:SER IBAS
DIAG:BOOT:WARM
DIAGnostic :BOOT[:WARM]
force boot
Range of
Values
SYSTem| IBASic| NONE
Note; :OWNer is implied
Complete the allocation
System Instrument Command Reference 7-7
Default
Units
none