Agilent Technologies E1300B User Manual page 136

75000 series b mainframes
Hide thumbs Also See for E1300B:
Table of Contents

Advertisement

DIAGnostic:COMMunicate :SERial[0][:OWNer]?
:COMMunicate
:SERial[0][:OWNer]?
Comments
Example
:COMMunicate
:SERial[n]:STORe
Comments
Example
7-8 System Instrument Command Reference
DIAGnostic:COMMunicate:SERial[0][:OWNer]? Returns the current "owner"
of the built-in serial interface. The values returned will be; "SYST", "IBAS", or
"NONE".
Related Commands: DIAGnostic:SERial[:OWNer]
Determine which instrument has the serial interface.
DIAG:COMM:SER?
enter statement
DIAGnostic:COMMunicate:SERial[n]:STORe Stores the serial
communications parameters (e.g. BAUD, BITS, PARity etc.) into non-volatile
storage for the serial interface specified by [n] in SERial[n].
Until
STORe is executed, communication parameter values are stored
in volatile memory, and a power failure will cause the settings to be lost.
DIAG:COMM:SER(1-7):STOR causes an Agilent E1324A (B-size
RS-232 card) to store its settings in an on-board EEROM. This EEROM
write cycle takes nearly one second to complete. Wait for this operation
to complete before attempting to use that serial interface.
The Agilent E1324A's EEROM used to store its serial communication
settings has a finite lifetime of approximately ten thousand write cycles.
Even if your application program sent the
every day, the lifetime of the EEROM would still be over 27 years.
Be careful that your application program sends the
to an Agilent E1324A no more often than is necessary.
Related Commands: all SYST:COMM:SER[n]
Store the serial communications settings in the third Agilent E1324A.
DIAG:COMM:SER3:STOR
Note; :OWNer is implied
statement returns the string
SYST, IBAS, or NONE
STORe command once
STORe command
commands