EA Technology
System Power – ON
Monitor Health – FLASHING during boot sequence, ON when monitoring.
Alarm and Outputs (1-3) – OFF
UltraBUS Power – OFF and consequently ON when in monitoring state
NOTE: When the UltraBUS Power LED is illuminated DO NOT connect/disconnect
any of the UltraBUS power connectors.
If the Hub Unit does not automatically start, press the Reset button. The boot sequence
should start immediately.
3. Alarms are configurable from the software by the end user. If any of the Alarm and Output
LEDs are ON, please analyse your settings in the software.
4. When the Monitor Health LED changes from Flashing to a steady state ON, the system will
be recording data.
5. To configure the Hub please refer to Section 11
The Monitor Hub may be shut down by one of the following two procedures.
1. If a remote connection can be established with the Monitor Hub then the procedures
detailed in the Software User Manual should be followed.
2. Alternatively the user may shutdown the Hub by pressing the Reset Button. The table below
details how this button can be used:
Hub Unit state
ON
Table 1: Reset button functionality
Health: This green LED will flash while the system is starting up, and will remain Power
Supply Operation
UltraBus: These sockets are for connection to the daisy chained Node communication bus.
These sockets are NOT compatible with Ethernet devices. Damage may be
caused to external Ethernet devices if they are plugged in to these sockets.
UltraTEV Monitor
Operating Manual
TM
Procedure
Short 1s press
Long 10s press
Long 10s press, 1s no press, 1s
press
13
Function
Sends the signal to shut down
and then auto restart the
Monitor. In order to power off
the monitor permanently, the
mains power must be
disconnected while the unit is
shutdown, before it restarts.
Once the Health LED goes out
completely, power can be
safely removed.
Force switch off. Apply only
if 1s press did not work.
Reset – Unit is forced to
switch off and then restarted.
Apply only if 1s press did
not work.
2178L014