Ecessary Appliances: Energy Saving Mode And Disconnecting Tube Adapter Including Bug-Fix - Otto Bock Genium X3 Setup Manual

Hide thumbs Also See for Genium X3:
Table of Contents

Advertisement

2 Necessary appliances: Energy saving
mode and disconnecting tube adapter
including bug-fix
Once the knee is fully charged, you'll need to perform the following sequences
of events before disconnecting and cutting the tube adapter to the correct length.
Start with the knee in full extension.
Flex the knee to 90 degrees.
Wait 5 seconds, power saving mode
Flex the knee until it reaches the flexion stop.
Disconnect the tube adapter.
If the knee head is fully flexed while the joint is standing upright, the joint switches
to sitting or energy-saving-mode after 2 seconds. Then the tube adapter can be
disconnected without triggering an error message. This function can be used for
storage, shipping or replacing the tube adapter.
If a knee error is detected, the knee will beep and/or vibrate repeatedly
Ensure the tube adapter is connected properly.
Check the battery status using the charger.
The knee will perform a self test.
If the knee beeps more than one time when the
charger is removed, contact the Ottobock service center.
will be activated.
Leave it fully flexed.
Connect the charger.
Disconnect the charger.
After disconnecting the charger the knee joint will
conduct a self test. Be sure not to move the knee joint
during this brief self test, since it can cause an error.
Genium X3 Set-up Guide | Ottobock 5