How to mark the adjustment range
Adjustment range: 30 mm total
40 mm
min.
insert
depth
70 mm max.
insert depth
How to roll the cable gently into the tube and insert the tube into the frame
The grid on the tube adapter must face anterior. Warning! Tube adapter insertion beyond 70 mm may cause damage to the
main electronics. When the Tube adapter is cut to the shortest length, please insert the blue plug into the recesses of the tube
adapter axial load sensor as seen in the photo below on the right side of the page.
12 Ottobock | Genium X3 Set-up Guide
How to connect the tube adapter
Caution: Do this gently!