ASBAR – Amputee Specific Bench Alignment Recommendations
Alignment and adjustment
Socket reference point
1  Position socket in a neutral position
2  Indicate the middle of the socket using the 743A80 (50-50 tool)
3  Draw a line through both points to the top of the socket
4  Find Ichial Tuberosity level and mark it on the line
5  Socket reference point = Ichial Tuberosity level + 30 mm
A personal alignment will be generated for the
user in X-Soft based on the information you en-
tered. The (P.R.O.S.A) assembly tool will lead to a
more accurate and reproducible alignment result.
Ensure the knee is in full extension during
bench alignment.
3 important reference points:
1    H ip rotation point 
Position in front of the bench alignment line
(the exact distance is displayed in X-soft)
2    K nee rotation point 
Position through the bench alignment line
(the exact distance is displayed in X-soft)
3    F oot middle 
Positioned in front of the bench alignment line
(the exact distance is displayed in X-soft)
With shoe
around
the foot.
Ensure the socket is not touching the hy-
draulic unit when the knee and/or adapters
are placed in maximum flexion.
Genium X3 Set-up Guide | Ottobock 15