5.5 Basic mode

Genium X3 created default settings that will be very close to
the physiologic settings of this user. When the need is there
this can be individualized by the:
CPO: Using X-soft in the whole range
User: Using the remote control in smaller range
Stance phase
Default setting is based on the user's body weight entered
in X-Soft
Adjust during sitting and optimize this setting by increas-
ing or decreasing stance resistance after the user has
walked down stairs and (steep) inclines.
Swing phase
Default setting is the physiologic swing angle: 65 degrees
Changing 1° can have a significant effect
Let the user walk allowing several steps to get used to this
function
Have the user walk at a slow, self selected and fast cadence
Functions
Train the user to use these functions.
We recommend "using" the Genium X3 Gait training DVD.
20 Ottobock | Genium X3 Set-up Guide