Beep and vibration signal
+
Type of error
Event
Warning
e. g. service interval has expired
Overheating of the hydraulic unit e. g. due
Hydraulic system
overheated
to uninterrupted, increased activity
e. g. a sensor (tube adapter, triple
sensor, ...) is not operating
Error
Selftest failure (if signals occur after
removing the charger)
AXON Tube Adapter has been discon-
nected while the knee joint is activated
and the inductive charger is not
AXON Tube Adapter
attached
disconnected
AXON Tube Adapter not connected
while removing inductive charger
e. g. failure of the valve drive
Critical error
Selftest failure (if signals occur after
removing the charger)
No functionality
No electronic functionality
Charging status 0%
Error code on the display
of the Remote Control
Using the Remote Control
+
+
+
+
+
10×
+
30 �
(periodically interrupted)
5× 
every minute
10×
continuous
5 minutes
intermittent
User restriction
Required action
No restriction
Consult Ottobock
Allow the hydraulic unit to cool down by
Reduced activity
reducing activity
Restart the knee joint by attaching and
removing inductive charger (up to three times);
If error remains, consult Ottobock immediately
Walking possible with
restrictions
Connect the AXON Tube Adapter
High stance resistance
Restart the knee joint by attaching and
Swing phase no longer
removing inductive charger (up to three times);
possible
If error remains, consult Ottobock immeditaly
Consult Ottobock immeditaly
Activation of safety
mode
Swing phase no longer
Charging required
possible
Consult Ottobock immeditaly (Error code
number required)
Select MyMode activated
Deep sleep mode activated
Charger removed, selftest OK
Charger removed; Warning
Error
AXON Tube Adapter
disconnected
Charging status 5%
Critical error
Genium X3 Set-up Guide | Ottobock 27