Auto System
Calibration
Thermo Fisher Scientific
25. Verify the 80i concentration reads similar to the 81i output
concentration value.
26. Record the following values:
Hg(0) Coefficient ___________________
Hg(t) Coefficient ___________________
Use the following procedure to perform an auto calibration.
1. From the Main Menu, select Instrument Controls > Service Mode.
2. Verify Service mode is OFF.
3. From the Main Menu, select Calibration > Auto Zero/Span Check.
4. Set the parameters shown in Table 3–1. Individual user requirements
may vary from the values shown in Table 3–1.
Table 3–1. Example of Parameter Settings
Parameter
Setting
next time
As Required
period hour
24 hours
inst zero durat min
10
inst span durat min
10
inst zero cal reset
ON
inst span cal reset
ON
inst hg span conc
Span 3
sys zero durat min
15
sys span durat min
15
sys zero cal reset
ON
sys span cal reset
ON
sys hg span conc
Span 3
purge duration min
0
zero/span avg sec
120
Calibration

Auto System Calibration

Mercury System Manual
3-3