Start-Up

Thermo Fisher Scientific
Example: From the Main Menu, select > Instrument Controls >
Gas Mode > Instrument Zero.
b. Verify that the 81i changes to that mode.
4. To test communication from the 80i to the 82i:
View Diagnostics from the 83i, such as Voltages > Interface Board
82i > Temperatures > Probe, or Pressure > Probe.
Verify that readings are not zero. Also, unplug one of the 83i
thermocouples and verify that the change is indicated on the 80i.
Use the following procedure to power up the Mercury CEMS.
1. Place the System in System Zero Mode to flush the system.
2. Turn OFF flow related components:
Power OFF the sample pump
Component Power > Eductor Power
Service > Probe Set Temp Failsafe > Set To (250 °C)
3. Heat components to temperatures:
Component Power > Umbilical 1 Power
Component Power > Umbilical 2 Power
Component Power > Converter Power
Component Power > Stinger Power
Component Power > Probe Power
Component Power > Oxidizer
4. Restart flow related items:
Service > Probe Set Temp Failsafe > Set To (normal Set-Point)
Component Power > Eductor Power
Power ON the sample pump
5. Place the system in Sample Mode.
System Installation, Set Up and Shut Down
Mercury System Manual
Start-Up
2-3