Thermo Scientific Mercury Freedom 80i Manual page 54

Table of Contents

Advertisement

Troubleshooting
System Troubleshooting Guides
4-8
Mercury System Manual
Malfunction
Instrument or system
appears to zero, but there is
weak or no response to
span gas.
Possible Cause
Action
less when the lamp is OFF, the
problem may be caused by scattered
light from dust in the optical bench.
Input board failure
Disconnect the input board from the
interface board by unplugging ribbon
cable labeled "INPUT." The instrument
reading should drop to zero or to a
negative value.
Critical orifice
Check/clean orifice.
blocked
External pump failure
Replace the external pump.
Insufficient air to 81i
Check the source pressure to the 81i.
The pressure should be 30 psig.
Calibration system
Check solenoids or other hardware to
failure
be sure that span gas is being
delivered to the correct spot.
Flow rate of the
Verify that the zero air system is
diluted span mix is
providing adequate flow.
inadequate
Critical orifice
Check/clean orifice.
blocked
Instrument is not
Check sample flow and pressure
drawing in span gas
readings (Diagnostics menu).
Use an independent flow meter to
check flows at the span inlet and
exhaust bulkheads (they should
match).
Perform a leak test.
Hg is being absorbed
Replace any lines made of vinyl or
by tubing, filters, or
other plastics with fresh Teflon.
dirt in the calibration
Replace Teflon filter membranes that
system
look dirty. Remove any filters that are
not Teflon membranes.
Lamp failed
Check the lamp intensity (Diagnostics
menu) and voltage.
81i output problem
Check 81i to verify output
concentration.
PMT or input board
With previous coefficients and PMT
has failed
voltage known, introduce a known
concentration of span gas.
Hydrator gas leak or
Check to see if bottom reservoir has
no water
water. Locate gas leak and repair.
Thermo Fisher Scientific

Advertisement

Table of Contents
loading

Table of Contents