Table 1-1. Model 81I Valve Mode Status - Thermo Scientific Mercury Freedom 80i Manual

Table of Contents

Advertisement

Table 1–1. Model 81i Valve Mode Status
Solenoid ID
Valve 1
Valve 2
Valve 3
Valve 4
Valve 5
Orifice Valve
Spike Valve
Thermo Fisher Scientific
(Model 82i), and a Hg probe (Model 83i), along with additional peripheral
components, such as a zero air supply, umbilical, and instrument rack.
However, the Model 81i is also available as a stand-alone calibrator.
Table 1–1 shows the Model 81i solenoid valve status for the different
calibrator modes. Use this table along with Figure 1–2 when reading the
principle of operation description that follows. Note that a valve status
shown as ON describes a valve that is energized, or in the Normally Closed
(NC) state. A valve status shown as OFF describes a valve that is de-
energized, or in the Normally Open (NO) state.
Standby
Instrument
Zero
ON
ON
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
Clean, dry, Hg-free pressurized air/nitrogen (30-40 psig) feeds both a high
flow (0-20 slpm) dilution air mass flow controller (MFC) and a low flow
(0-50 sccm) Hg source MFC (Figure 1–2).
Instrument
Orifice
Cal
Zero
ON
ON
ON
OFF
ON
OFF
OFF
ON
ON
ON
OFF
ON
OFF
OFF
Orifice
System
System
Cal
Zero
Cal
ON
ON
ON
ON
OFF
ON
ON
OFF
ON
OFF
ON
OFF
OFF
ON
OFF
ON
OFF
OFF
OFF
ON
ON
Mercury System Manual
Introduction
Model 81i
1-3

Advertisement

Table of Contents
loading

Table of Contents