Plumbing And Electrical Hookup; Set Temperature Defaults; Mercury Cems Power-Up Sequence; Table 2-1. Default Temperatures - Thermo Scientific Mercury Freedom 80i Manual

Table of Contents

Advertisement

System Installation, Set Up and Shut Down

Plumbing and Electrical Hookup

Plumbing and
Electrical Hookup
Set Temperature
Mercury CEMS
Power-Up Sequence
2-2
Mercury System Manual
Refer to the "System Reference Drawings" in Appendix B for the system
plumbing and electrical hookup connections.
Equipment Damage To prevent damaging the probe transducer, read the
cable labels carefully and DO NOT connect the C or D cable to the wrong
connector on the 82i rear panel.
Refer to Table 2–1 for setting the default temperatures (or per site
requirements). At the 80i Main Menu, select > Service > Set
Defaults
Temperatures:
Table 2–1. Default Temperatures
Parameter
Bench Temp
Probe Temp
Umbilical Temp
Converter Temp
Failsafe Valve Temp
Oxidizer Temp
Use the following procedure to power up the Mercury CEMS.
Note If the 81i is in Service Mode, it will ignore 80i commands.
1. Apply Power to the CEMS, turn power switches ON in the following
order:
2. After the Model 81i initializes and the Run screen is displayed, power
the Model 80i.
3. To test communication from the 80i to the 81i:
a. Using the Model 80i, change the Gas Mode to one of the
Default Temperature
Setting
45 °C
220 °C
120 °C
760 °C
200 °C
400 °C
Model 82i
Model 81i
following: Instrument Zero, Instrument Cal, System Zero, or
System Cal.
Thermo Fisher Scientific