Tuning and Calibrating

Running Auto Tune and Calibration

Figure 32. EZ Tune - EZ Workspace, showing the FC-43 mass spectrum in EI, positive polarity mode before tuning and
calibrating.
Running Auto Tune and Calibration
66
TSQ Quantum XLS and TSQ Quantum GC User Guide
Run the tune and calibration procedure after you obtain a good FC-43 ion signal.
To tune and calibrate your mass spectrometer automatically in the EI, positive ion
mode
1. Display the FC-43 positive ion mass spectrum as described in the previous section.
2. Click the System Tune and Calibration button (or choose Setup > System Tune and
Calibration) to display the System Tune and Calibration dialog box.
Thermo Scientific