Alarm Details and Actions

Alarm
Elevated Pressure
(2cm H
O over Set
2
Value)
Low Pressure
(2cm H
O under
2
Set Value)
High Pressure
(≥12 cm H
O for 2
2
seconds or
≥18 cm H
O
2
instantaneously)
Disconnect
(Table Based)
% O
Alarms
2
(Utilizes an oxygen
time product. Time
product is defined
as: the variance
from FiO
set point
2
multiplied by the
amount of time of
the variance.)
®
AirLife
nCPAP Operator's Manual
Priority
Medium
• Audible alarm
• Yellow flashing LED
• Message Bar
Medium
• Audible alarm
• Yellow flashing LED
• Message Bar
High
• Audible alarm
• Red flashing LED
• Message Bar
• The patient delivery
valve closes and a
relief valve opens
immediately upon
alarm. System
attempts to reinitiate
treatment every three
(3) seconds until the
alarm condition clears.
Medium
• Audible alarm
• Yellow flashing LED
• Message Bar
• The flow rate is
adjusted to Target
Flow ±2 LPM.
Medium
• Audible alarm
(20 %O
*
• Yellow flashing LED
2
sec)
• Message Bar
High
• Audible alarm
(30 %O
*
• Red flashing LED
2
sec)
• Message Bar.
Indication
Suggested Action
• Check the exhaust tube
for occlusions or
restrictions
• Check for leaks in the
tubes and circuit, or in the
nasal prongs and mask
• Check supply gasses
• Ensure that Auxiliary
flow is not set higher than
the values in Table 1.
• Check the exhaust tube
for occlusions or
restrictions
• Check for leaks or
disconnections in the
tubes and circuit, or in the
nasal prongs and mask.
• Check Supply Gasses
• Restart driver to perform
oxygen sensor calibration
• Check supply gasses
• Restart driver to perform
oxygen sensor calibration
23