Cardinal Health AirLife 006900 Operator's Maintenance Manual page 26

Ncpap system
Table of Contents

Advertisement

Alarm
Leak
(3 LPM over the
Target Flow rate
or a Flow Rate
of >15 LPM)
Low Flow
(At or below 3
LPM flow rate)
External Power
Disconnect
Battery Low
Dual Stage Alarm
Battery Low
Dual Stage Alarm
Battery Charge
Fault
Oxygen Pressure
Low
Oxygen Pressure
High
24
Priority
Medium
• Audible alarm
• Yellow flashing LED
• Message Bar
High
• Audible alarm
• Red flashing LED
• Message Bar
Low
• Single audible burst
• Indicator on display
• LED turns off
Medium
• Audible alarm
(~24 minutes
• Yellow LED
remaining)
High
• Audible alarm
(~12 minutes
• Red Flashing LED
remaining)
• Message Bar
High
• Single Audible burst
• Red Power LED
• LCD indicator
High
• Audible alarm
• Red Flashing LED
• Message Bar
High
• Audible alarm
• Red Flashing LED
• Message Bar
Indication
®
AirLife
nCPAP Operator's Manual
Suggested Action
• Check for leaks or
disconnections in the
tubes and circuit, or in the
nasal prongs and mask
• If leak can not be
corrected by re-
positioning or re-sizing the
patient interface, the
Target flow can be
increased to allow the
driver to deliver higher
flows when necessary.
• Check the exhaust tube
for occlusions or
restrictions
• Plug in external power
• Plug in external power
• Plug in external power
• System will be required
to have external power
connected to ensure
patient treatment is not
interrupted.
• Correct oxygen supply
pressure
• Select 'Air Only' mode
• Follow on-screen
directions
• Correct oxygen supply
pressure
• Select 'Air Only' mode
• Follow on-screen
directions