Appendix B: Electromagnetic Compatibility

The following information is directed by the Standard IEC 60601-1-2:2001
for Medical Electrical Equipment - Part 1-2: General Requirements for Safety
- Collateral Standard: Electromagnetic Compatibility - Requirements and
Tests.
Warning: The use of any power cords with the AirLife
nCPAP System Driver other than those supplied with the
driver may result in increased emissions and/or decreased
Immunity of the AirLife
Table 2: Guidance and Manufacturer's Declaration—Electromagnetic
®
The AirLife
nCPAP System Driver is intended for use in the electromagnetic
environment specified below. The customer or user of the AirLife
should assure that it is used in such an environment.
Emissions Test
CISPR 11
RF Emissions
CISPR 11
RF Emissions
IEC 61000-3-2
Harmonic
Emissions
IEC 61000-3-3
Voltage
Fluctuations,
Flicker Emissions
30
®
nCPAP System Driver.
Emissions
Compliance
The AirLife
energy only for its internal function. RF emissions
Group 1
are very low and are not likely to cause any
interference in nearby electronic equipment.
Class B
The AirLife
Not
use in all establishments, including domestic
Applicable
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Not
Applicable
Electromagnetic Environment—Guidance
®
nCPAP System Driver uses RF
®
nCPAP System Driver is suitable for
®
AirLife
nCPAP Operator's Manual
®
®
nCPAP System Driver