Alarm
Air Pressure Low
Air Pressure High High
Oxygen Sensor
Calibration Error
Flow / Pressure
Sensor Failure
Software Fault
®
AirLife
nCPAP Operator's Manual
Priority
High
• Audible alarm
• Red Flashing LED
• Message Bar
• Audible alarm
• Red Flashing LED
• Message Bar
High
• Audible alarm
• Red Flashing LED
• Calibration Error
Screen
High
• All LEDs illuminated
• Sensor Calibration
Failure screen
High
• Audible alarm
• Red Flashing LED
• Message Bar
Indication
Suggested Action
• Correct air supply
pressure
• Select 'Oxygen Only'
mode
• Follow on-screen
directions
• Correct air supply
pressure
• Select 'Oxygen Only'
mode
• Follow on-screen
directions
• Follow on-screen
directions
• While alarm is active,
the driver provides CPAP
and FiO
; however, the
2
FiO
may not be accurate.
2
• Remove patient from
device and restart driver
• Make sure Patient is
disconnected from circuit
and restart driver.
• Follow on-screen
directions
• Record the Fault code
• Follow on-screen
directions
25