Cardinal Health AirLife 006900 Operator's Maintenance Manual page 33

Ncpap system
Table of Contents

Advertisement

Table 3: Guidance and Manufacturer's Declaration—Electromagnetic Immunity
®
The AirLife
nCPAP System Driver is intended for use in the electromagnetic environment
specified below. The customer or user of the AirLife
that it is used in such an environment.
Immunity Test
IEC 61000-4-2
Electro-Static
Discharge (ESD)
IEC 61000-4-4
Electrical Fast
Transient, Burst
(EFT) (Note 2)
IEC 61000-4-5
Power Line Surge
(Note 2)
IEC 61000-4-8
Power Frequency
Magnetic Field
(50/60 Hz)
IEC 61000-4-11
Voltage Dips,
Short
Interruptions, and
Voltage Variations
(Note 2)
Note 1—UT is the AC mains voltage prior to application of the test level.
Note 2—Performed with an Input Voltage of 120 V AC only.
®
AirLife
nCPAP Operator's Manual
IEC 60601-1-2
Test Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV common
mode
3 A/m
<5 % UT
(Note 1)
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
®
nCPAP System Driver should assure
Compliance
Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV common
mode
3 A/m
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Electromagnetic
Environment—Guidance
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
If the user of the AirLife
nCPAP System Driver
requires continued operation
during power mains
interruptions, it is
recommended that the
®
AirLife
nCPAP System
Driver be powered from a
non-interruptible power
supply or a battery.
®
31