6.
Verify that the following default values are indicated on the main screen:
Note: Alarms are delayed for 120 seconds after entering the
main screen in order to allow time for patient setup. If the
patient setup takes longer than 120 seconds, the alarms can
be muted for an additional 120 seconds by pressing the alarm
mute/reset button or the system can be placed into Standby
Mode. See Figure 8.
7.
Unplug external power, verify a single audible alarm burst sounds and
the interface indicates the driver is operating on battery power. Plug
external power source back in.
Caution: The power switch does not disconnect the driver from
the main power supply. This can only be accomplished by
disconnecting the AC adapter from either the driver or the AC
outlet.
®
AirLife
nCPAP Operator's Manual
Figure 7: Illuminated LEDs
CPAP Set: 5 cmH
FiO
Set: 21% O
2
Flow Rate Target: 8 LPM
Battery charging (lightning bolt in battery icon)
Figure 8: Alarm Mute/Reset Button
O
2
2
15