Performance Tests
Table 1–12: Bessel Thompson frequency response and reference receiver limits
Standard
Frequency (MHz)
OC–12 SONET or STM–4 SDH, Data Rate: 622.08 Mb/s
0.0
93.31
186.6
279.9
373.2
466.7
5.60.0
622.08
653.2
746.5
839.8
933.1
1244.2
OC–3 SONET or STM–1 SDH, Data Rate: 155.52 Mb/s
0.0
23.327
46.65
69.975
93.3
116.88
139.97
155.52
163.3
209.95
209.95
233.27
311.04
1–100
Lower Limit (dB)
Nominal (dB)
–0.3
0.0
–0.41
–0.11
–0.75
–0.45
–1.32
–1.02
–2.16
–1.86
–3.31
–3.00
–5.15
–4.51
–6.58
–5.71
–7.35
–6.37
–9.86
–8.54
–12.59
–10.93
–15.41
–13.41
–24.58
–21.44
–0.3
–0.0
–0.41
–0.11
–0.75
–0.45
–1.32
–1.02
–2.16
–1.86
–3.31
–3.00
–5.15
–4.51
–6.58
–5.71
–7.35
–6.37
–9.86
–8.54
–12.59
–10.93
–15.41
–13.41
–24.58
–21.44
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Upper Limit (dB)
Measured (delta dB)
0.3
0.19
–0.15
–0.72
–1.56
–2.71
–3.87
–4.84
–5.39
–7.22
–9.27
–11.41
–18.31
0.3
0.19
–0.15
–0.72
–1.56
–2.71
–3.87
–4.84
–5.39
–7.22
–9.27
–11.41
–18.31