TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Use the keypad to set the horizontal scale (/div) to 5 ms (press 5,
SHIFT, m, then ENTER).
b. Set up 60 Hz signal generator:
Connect the output of the signal generator to the CH2 input through
a 50
cable. See Figure 1–31.
Signal Generator
Output
Figure 1–31: 60 Hz Rejection test hookup
Adjust the signal generator for three vertical divisions of 60 Hz
signal. See Figure 1–32. The signal will not be triggered. That is, it
will run free.
Figure 1–32: 60 Hz Rejection test setup signal
Oscilloscope
50
cable
Performance Tests
1–77