Figure 1-24: Subsequent Test Hookup - Tektronix TDS 700D Technical Reference

Digitizing oscilloscopes
Hide thumbs Also See for TDS 700D:
Table of Contents

Advertisement

TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Connect the Sense output of the generator through a second
dual-banana connector followed by a 50
the other side of the BNC T connector. Now connect the BNC T
connector to CH 1. See Figure 1–24.
DC Calibrator
Dual banana to
BNC adapters
Figure 1–24: Subsequent test hookup
e. Measure amplitude of the probe compensation signal:
Press SHIFT; then press ACQUIRE MENU. Press the side-menu
button AVERAGE then enter 16 using the keypad or the general
purpose knob.
Adjust the output of the DC calibration generator until it precisely
overlaps the top (upper) level of the stored probe compensation
signal. (This value will be near 500 mV).
Record the setting of the DC generator.
Adjust the output of the DC calibration generator until it precisely
overlaps the base (lower) level of the stored probe compensation
signal. (This value will be near zero volts).
Record the setting of the DC generator.
f.
Press CLEAR MENU to remove the menus from the display. See
Figure 1–25.
precision coaxial cable to
50
coaxial cables
Performance Tests
Oscilloscope
BNC T
connector
1–69