Table 1-4: Analog Bandwidth - Tektronix TDS 700D Technical Reference

Digitizing oscilloscopes
Hide thumbs Also See for TDS 700D:
Table of Contents

Advertisement

Table 1–4: Analog bandwidth
Reference amplitude
Re ere ce ampli
Vertical scale
er ical cale
(6 divisions)
100 mV
600 mV
1 mV
6 mV
2 mV
12 mV
5 mV
30 mV
10 mV
60 mV
20 mV
120 mV
50 mV
300 mV
200 mV
1.2 V
1
500 mV
3 V
1
1 V
6 V
1
If your generator cannot output 6 divisions of amplitude, set it to its maximum output. The –3 db limit can be calculated as:
0.707
reference amplitude.
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
e
Horizontal
ri
al
scale
TDS 794D
1 ns
2 GHz
1 ns
NA
1 ns
NA
1 ns
NA
1 ns
2 GHz
1 ns
2 GHz
1 ns
2 GHz
1 ns
2 GHz
1 ns
2 GHz
1 ns
2 GHz
e. Display the test signal: Do the following subparts to first display the
reference signal and then the test signal.
Press MEASURE; then press the main-menu button Select
Measrmnt for CHx.
Press the side-menu button more, if needed, until the menu label
Frequency appears in the side menu (its icon is shown at the left).
Press the side-menu button Frequency.
Press the side-menu button more until the menu label Pk-Pk
appears in the side menu (its icon is shown at the left). Press the
side-menu button Pk-Pk.
Press CLEAR MENU.
Set the generator output so the CHx Pk-Pk readout equals the
reference amplitude in Table 1–4 that corresponds to the vertical
scale set in substep d.
Press the front-panel button SET LEVEL TO 50% as necessary to
trigger a stable display. At full bandwidth, you may also want to
make small, manual adjustments to the trigger level. You can use the
TRIGGER LEVEL knob to do this. (Full bandwidth varies with
TDS model as is shown in Table 1–4).
Test frequency
TDS 580D
TDS 680B
TDS 620B
TDS 684B
TDS 644B
TDS 784D
1 GHz
500 MHz
500 MHz
250 MHz
600 MHz
300 MHz
750 MHz
450 MHz
1 GHz
500 MHz
1 GHz
500 MHz
1 GHz
500 MHz
1 GHz
500 MHz
1 GHz
500 MHz
1 GHz
500 MHz
Performance Tests
TDS 520D
TDS 540D
TDS 724D
–3 b Limi
–3 db Limits
TDS 754D
500 MHz
424 mV
450 MHz
4.24 mV
500 MHz
8.48 mV
500 MHz
21.2 mV
500 MHz
42.4 mV
500 MHz
84.8 mV
500 MHz
212 mV
500 MHz
848 mV
1
500 MHz
2.12 V
1
500 MHz
4.24 V
1–39