Performance Tests
1–70
Figure 1–25: Measurement of probe compensator amplitude
g. Check against limits:
Subtract the value just obtained (base level) from that obtained
previously (top level).
CHECK that the difference obtained is within 495 mV to 505 mV,
inclusive.
Enter voltage difference on test record.
3. Disconnect the hookup: Disconnect the cable from CH 1.