Performance Tests
Read results.
Set the generator
(reference) frequency to
the test frequency from
Table 1–4.
Set the horizontal scale
from Table 1–4.
1–40
f.
Measure the test signal:
Set the frequency of the generator, as shown on screen, to the test
frequency in Table 1–4 that corresponds to the vertical scale set in
substep d. See Figure 1–9.
Set the horizontal SCALE to the horizontal scale setting in
Table 1–4 that corresponds to the vertical scale set in substep d.
Press SET LEVEL TO 50% as necessary to trigger the signal.
Read the results at the CHx Pk-Pk readout, which will automatically
measure the amplitude of the test signal. See Figure 1–9.
3
1
2
Figure 1–9: Measurement of analog bandwidth
g. Check against limits:
CHECK that the Pk-Pk readout on screen is within the limits listed
in Table 1–4 for the current vertical scale setting.
Enter voltage on test record.
When finished checking, set the horizontal SCALE back to the
50 ns setting.
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications