Performance Tests

Sine Wave Generator Leveling Procedure

1–88
Some procedures in this manual require a sine wave generator to produce the
necessary test signals. If you do not have a leveled sine wave generator, use one
of the following procedures to level the output amplitude of your sine
wave generator.
Equipment
Sine wave generator (Item 14)
required
Level meter and power sensor (Item 15)
Power splitter (Item 16)
Two male N to female BNC adapters (Item 17)
One precision coaxial cable (Item 5)
Prerequisites
See page 1–15
Sine Wave
Generator
Power splitter
Output
Power sensor
Figure 1–41: Sine wave generator leveling equipment setup
1. Install the test hookup: Connect the equipment as shown in Figure 1–41.
2. Set the Generator:
Set the sine wave generator to a reference frequency of 10 MHz.
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications