Figure 1-5: Initial Test Hookup - Tektronix TDS 700D Technical Reference

Digitizing oscilloscopes
Hide thumbs Also See for TDS 700D:
Table of Contents

Advertisement

Performance Tests
Check DC Voltage
Measurement Accuracy
1–32
WARNING. The generator is capable of outputting dangerous voltages. Be sure to
set the DC calibration generator to 0 volts before connecting, disconnecting,
and/or moving the test hookup during the performance of this procedure. Also,
check that the calibrator does not have shorting straps installed between the DC
and sense outputs or grounds.
Equipment
Two dual-banana connectors (Item 6)
required
One BNC T connector (Item 7)
One DC calibration generator (Item 9)
Two precision 50
Prerequisites
The oscilloscope must meet the prerequisites listed on page 1–15
See Input Channels vs. Model on page 1–2.
1. Install the test hookup and preset the instrument controls:
a. Hook up the test-signal source:
Set the output of a DC calibration generator to 0 volts.
Connect the output of a DC calibration generator through a
dual-banana connector followed by a 50
one side of a BNC T connector. See Figure 1–5.
Connect the Sense output of the generator through a second
dual-banana connector followed by a 50
the other side of the BNC T connector. Now connect the BNC T
connector to CH 1. See Figure 1–5.
Output Sense
DC Calibrator
Dual banana to
BNC adapters
Figure 1–5: Initial test hookup
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
coaxial cables (Item 5)
precision coaxial cable to
precision coaxial cable to
50
coaxial cables
Oscilloscope
BNC T
connector

Advertisement

Table of Contents
loading

Table of Contents