Figure 1-16: Initial Test Hookup - Tektronix TDS 700D Technical Reference

Digitizing oscilloscopes
Hide thumbs Also See for TDS 700D:
Table of Contents

Advertisement

Check Accuracy,
Trigger-Level or
Threshold, DC Coupled
Dual banana to
BNC adapters
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Equipment
One DC calibration generator (Item 9)
required
One BNC T connector (Item 7)
Two 50
Prerequisites
See page 1–15.
DC Calibrator
50
coaxial cables
Figure 1–16: Initial test hookup
1. Install the test hookup and preset the instrument controls:
a. Hook up the test-signal source:
Set the output of the DC calibration generator to 0 volts.
Connect the output of the DC calibration generator, through a
dual-banana connector followed by a 50
to one side of a BNC T connector.
Connect the Sense output of the generator, through a second
dual-banana connector followed by a 50
to other side of the BNC T connector. Now connect the BNC T
connector to CH 1.
b. Initialize the oscilloscope:
Press save/recall Setup.
Press the main-menu button Recall Factory Setup.
Press the side-menu button OK Confirm Factory Init.
2. Confirm Main trigger system is within limits for Trigger-level/Threshold
accuracy:
a. Display the test signal:
precision coaxial cables (Item 5)
Oscilloscope
BNC T
connector
Performance Tests
precision coaxial cable,
precision coaxial cable,
1–53