Figure 1-30: Triggered Signal Range Test - 75 Mv - Tektronix TDS 700D Technical Reference

Digitizing oscilloscopes
Hide thumbs Also See for TDS 700D:
Table of Contents

Advertisement

Performance Tests
Positive pulse
Negative pulses
1–76
n. Use the keypad to set the fine scale to 75 mV (press 75, SHIFT, m, then
ENTER).
o. CONFIRM that the TRIG'D LED stays lighted and that the waveform
on screen is stable. That is, it does not move horizontally or vertically.
Also, CONFIRM that the waveform on the screen has one positive pulse
and a number of negative pulses. See Figure 1–30.
Figure 1–30: Triggered signal range test – 75 mV
p. Disconnect all test equipment (TSG121) from the oscilloscope.
5. Check 60 Hz Rejection.
a. Set up oscilloscope for 60 Hz Rejection Test:
Use the keypad to set the Ch1 Fine Scale to 282 mV (press 282,
SHIFT m, then ENTER).
Press WAVEFORM OFF.
Press CH2.
Press VERTICAL MENU.
Use the keypad set the fine scale to 2 V (press 2, then ENTER).
Press HORIZONTAL MENU.
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications

Advertisement

Table of Contents
loading

Table of Contents