Figure 1-6: Measurement Of Dc Accuracy At Maximum Offset And Position - Tektronix TDS 700D Technical Reference

Digitizing oscilloscopes
Hide thumbs Also See for TDS 700D:
Table of Contents

Advertisement

Turn on the
measurement called
mean and read the
results here.
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Figure 1–6: Measurement of DC accuracy at maximum offset and position
f.
Check against limits:
CHECK that the readout for the measurement Mean readout on
screen is within the limits listed for the current vertical scale and
position/offset/generator settings. Enter value on test record.
Repeat substep d, reversing the polarity of the position, offset, and
generator settings as is listed in the table.
CHECK that the Mean measurement readout on screen is within the
limits listed for the current vertical scale setting and position/offset/
generator settings. Enter value on test record.
Repeat substeps c through f until all vertical scale settings, listed in
Table 1–3, are checked for the channel under test.
g. Test all channels: Repeat substeps a through f for all four channels.
3. Disconnect the hookup:
a. Set the generator output to 0 V.
b. Disconnect the cable from the generator output at the input connector of
the channel last tested.
Performance Tests