Output Signal Checks; Figure 1-20: Initial Test Hookup - Tektronix TDS 700D Technical Reference

Digitizing oscilloscopes
Hide thumbs Also See for TDS 700D:
Table of Contents

Advertisement

Performance Tests

Output Signal Checks

Check Outputs — CH 3
(AUX 1 on some models)
Main and Delayed Trigger
1–62
The procedure that follows checks those characteristics of the output signals that
are listed as checked under Warranted Characteristics in Specifications. The
oscilloscope outputs these signals at its front and rear panels.
Equipment
Two precision 50
required
One calibration generator (Item 10)
Prerequisites
See page 1–15. Also, the oscilloscope must have passed Check DC
Voltage Measurement Accuracy on page 1–32.
See Input Channels vs. Model on page 1–2.
1. Install the test hookup and preset the instrument controls:
Calibration
To MAIN
Generator
TRIGGER
OUT
Figure 1–20: Initial test hookup
a. Hook up test-signal source 1:
Connect the standard amplitude output of a calibration generator
through a 50
TDS models).
Set the output of the calibration generator to 0.500 V.
b. Hook up test-signal source 2: Connect the Main Trigger Out at the rear
panel to CH 2 through a 50
c. Initialize the oscilloscope:
Press save/recall SETUP.
Press the main-menu button Recall Factory Setup.
Press the side-menu button OK Confirm Factory Init.
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
coaxial cables (Item 5)
Oscilloscope
50
coaxial cables
precision coaxial cable to CH 3 (AUX1 on some
precision cable.