Performance Tests
1–56
d. Check against limits: Do the following subparts in the order listed.
CHECK that the Level readout in the side menu is within the limits
listed in Table 1–7.
Table 1–7: Delayed trigger level accuracy readout
Instrument
TDS 600B
TDS 500D/700D
TDS 794D
Enter voltage on test record.
Press the main-menu button Slope; then press the side-menu button
for negative slope. See icon at left. Press the main-menu button
Level. Repeat substep c.
CHECK that the Level readout in the side menu is within the limits
listed in Table 1–7.
Enter voltage on test record.
4. Disconnect the hookup:
a. First set the output of the DC calibration generator to 0 volts.
b. Disconnect the cable from the generator output at the input connector of
CH 1.
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Minimum
9.863 V
9.8853 V
4.8953 V
Maximum
10.137 V
10.1147 V
5.1047 V