Rigol MSQ7054 User Manual page 153

Digital oscilloscope
Table of Contents

Advertisement

Chapter 6 Operations and Measurements
Table 6-1 Window Function
Window
Function
Rectangular
Blackman-Harris
Hanning
Hamming
Flattop
Triangle
For FFT math operation setting label, refer to descriptions in "Math Operation
Label".
Press More  More, and press Color Grade continuously to enable or disable
the color grade display of the FFT operation results. When enabled, different
colors are displayed on the screen to indicate the times of data acquisition or
acquisition probability.
Press More  More  Reset ColorGrade to clear the color grade display and
display the color grade again.
MSO7000/DS7000 User Guide
Characteristics
Best frequency
resolution
Poorest
amplitude
resolution
Similar to the
situation when
no window is
applied.
Best amplitude
resolution
Poorest
frequency
resolution
Better frequency
resolution and
poorer amplitude
resolution
compared with
Rectangular
A little bit better
frequency
resolution than
Hanning
Measure the
signals
accurately
Better frequency
resolution
Waveforms Applicable to the
Window Function
Transient or short pulse, the signal
levels before and after the
multiplication are basically the same.
Sine waveforms with the same
amplitudes and rather similar
frequencies.
Wide band random noise with
relatively slow change of waveform
spectrum.
Single frequency signal, searching for
higher order harmonics.
Sine, periodic, and narrow band
random noise.
Transient or short pulse, the signal
levels before and after the
multiplication are rather different.
Measure the signal that has no
accurate reference and requires an
accurate measurement
Measure the narrow band signal and
that has strong noise interference
RIGOL
6-9

Advertisement

Table of Contents
loading