Emerson Rosemount 2160 Reference Manual page 64

Wireless level detector
Hide thumbs Also See for Rosemount 2160:
Table of Contents

Advertisement

Service and troubleshooting
May 2022
If the difference is greater than Allowable Change In Dry Fork Frequency, the re-calibration
is rejected. Check the fork for damage, corrosion, or coating, and clean it if necessary
before re-trying.
When the calibration is successful, Dry Fork Frequency is set to the new dry frequency.
Related information
Allowable change in fork frequency
Start the dry fork calibration
Procedure
1. Select Service Tools → Maintenance → Calibrate.
2. Select Calibrate Dry Fork and follow the on-screen instructions to perform the dry
Sensor calibration status
Sensor Calibration Status indicates:
Status
Factory calibrated
Site calibrated
Un-calibrated
Restoring factory calibration of dry fork sensor
This command restores the factory calibration of the fork sensor in dry process conditions.
It is accessible after a site calibration has been performed, but should only be performed
by authorized persons.
Restoring site calibration of dry fork sensor
This command restores the site calibration of the fork sensor in dry process conditions. It is
accessible after the restoration of a factory calibration has been performed, but should
only be performed by authorized persons.
6.4.5
Adjust sensor compensation
The frequency of the vibrating fork sensor may be affected by a process temperature
being different to the calibrated temperature. Entering a known process temperature can
compensate accordingly and provide an improved frequency switching point.
Procedure
1. Select Service Tools → Maintenance → Calibrate.
2. Under Sensor Compensation, enter the known Process Temperature.
64
fork calibration.
Description
No further calibration is normally required.
Calibration successfully performed on-site.
Calibration is required. Contact the factory.
Reference Manual
00809-0100-4160
Reference Manual

Advertisement

Table of Contents
loading

Table of Contents