Apmi Modules - Ametek Crystal HPC51 Operation Manual

Intrinsically safe handheld pressure calibrator
Table of Contents

Advertisement

APMi Modules

       
ACCURACY
psi (Gauge Pressure)
18 to 28° C
X
0 to 30% of Range: . . . . . . . . . . . . . . . ±(0.01% of Full Scale)
30 to 110% of Range: . . . . . . . . . . . . ±(0.035% of Reading)
*
Vacuum
: . . . . . . . . . . . . . . . . . . . . . . . ±(0.05% of Full Scale
-20 to 50° C
X
0 to 30% of Range: . . . . . . . . . . . . . . . ±(0.015% of Full Scale)
30 to 110% of Range: . . . . . . . . . . . . ±(0.050% of Reading)
*
Vacuum
: . . . . . . . . . . . . . . . . . . . . . . . ±(0.05% of Full Scale
*
Applies to 300 psi / 30 bar / 3 MPa and lower ranges only .
Vacuum Range = -14 . 5 psi / -1 . 0 bar / -1MPa .
**
Full Scale is the numerical value of the positive pressure range .
psiA (Absolute Pressure with BARO Option)
All absolute accuracies are equivalent to the gauge pressure
accuracies, except as noted below.
15 psi / 1 bar / 100 kPa Range: . . . . . Gauge Accuracy + 0.005 psiA
30 psi / 3 bar / 300 kPa Range: . . . . . Gauge Accuracy + 0.005 psiA
100 psi / 10 bar / 1MPa Range: . . . . . Gauge Accuracy + 0.002 psiA
Includes all effects of linearity, hysteresis, repeatability, temperature,
and stability for one year .
Exposure to environmental extremes of temperature, shock, and/or
vibration may warrant a more frequent recertification period .
APMi modules must be exercised and re-zeroed whenever exposed
to significant changes in environmental conditions to achieve these
specifications . To exercise a module, cycle the module between zero
(ambient barometric pressure) and the pressure of interest . A properly
exercised module will return to a zero reading (or return to the same
ambient barometric reading) .
       
ACCURACY CONTINUED
All models indicate vacuum, but vacuum specification applies only
where specified .
Not recommended for continuous use at high vacuum. Refer to
XP2i-DP data sheet
for gauges that are intended for continuous high
vacuum use .
The BARO option for the HPC50 Series allows you to toggle between
* *
)
gauge and absolute pressure using the same APMi module .
       
SENSOR
Wetted Materials. . . . . . . . . . . . .(wrench tight) 316 stainless steel
* *
)
Diaphragm Seal Fluid . . . . . . . .Silicone Oil
Connection . . . . . . . . . . . . . . . . . .Crystal CPF Female
All welded, with a permanently filled diaphragm seal .
Metal to metal cone seal; O-ring can be removed if necessary .
1/4" male NPT adapter is included unless BSP or M20 is specified .
1/4" medium pressure tube system is compatible with HIP LM4 and
LF4 Series, Autoclave Engr SF250CX Male and Female Series .
       
OUTPUT
Pressure Resolution . . . . . . . . . .up to 6 digits
Display Update . . . . . . . . . . . . . .up to 10 per second
Pressure resolution and display update are the maximum values
available . The resolution of your Crystal device may be different .
       
PRESSURE OVERLOAD
Overload Alarm . . . . . . . . . . . . . ."+OL" in display at 110% F.S.
       
OPERATING TEMPERATURE
Temperature Range . . . . . . . . . .-20 to 50° C (-4 to 122° F)
< 95% RH, non-condensing . No change in accuracy over operating
temperature range, except as noted in the accuracy specifications .
APMi must be zeroed to achieve rated specification .
Applies to all modules .
       
STORAGE TEMPERATURE
(finger tight) 316 stainless steel
and Viton® with internal o-ring
Temperature Range . . . . . . . . . .-40 to 75° C (-40 to 167° F)
(15 psi / 1 bar / 100 kPa) 316 stain-
       
ENCLOSURE
less steel and Viton®
Dimensions . . . . . . . . . . . . . . . . . .2.9 x 1.1 in (74.0 x 27.0 mm)
Weight . . . . . . . . . . . . . . . . . . . . . .0.39 lbs (176.0 g)
7/16-20 MP CPF PORT
APMi Modules
21
STAINLESS STEEL HOUSING
Ø 27
(1.06)
74.0
(2.92)
6 PIN CONNECTION
FOR HPC50 SERIES CABLE
HPC50 Series Operation Manual