Ametek Crystal Pressure XP2i-DD Operation Manual

Ametek Crystal Pressure XP2i-DD Operation Manual

Digital test gauge (standard and dual-display(dd))

Advertisement

Quick Links

XP 2 i
Operation Manual
for XP2i Digital Test Gauge (Standard and Dual-Display(DD))

Advertisement

Table of Contents
loading

Summary of Contents for Ametek Crystal Pressure XP2i-DD

  • Page 1 XP 2 i Operation Manual for XP2i Digital Test Gauge (Standard and Dual-Display(DD))
  • Page 2: Table Of Contents

    Contents Overview Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
  • Page 3: Overview

    We hope your XP2i meets your expectations, and we're interested in any comments or suggestions you may have. You can send us a note at: A user defined pressure scale, and/or disable crystal@ametek.com. Many features in this and our other products are a direct result of your comments! unused pressure units •...
  • Page 4: Operating Instructions

    Overview  „ „ „ „ „ OPERATING INSTRUCTIONS The XP2i is shipped with batteries installed, so it’s ready to use. Press and hold the on/off button. The XP2i will first test all LCD segments. Release on/off button when the XP2i indicates pressure. The XP2i always resumes operation in the mode and the units of the pressure last used, and it does not automatically rezero when turned on.
  • Page 5: Functions

    Functions  Functions Units Button units Pressing this button causes the XP2i to select the next available unit of pressure measurement. Pressure Ranges, Display Scales, & Resolution on page 29 or the list of pressure units available for your model. Units that you don’t need or never use can be turned off.
  • Page 6 Functions  zero Resetting (Clearing) Recorded Peak Values clear Peak values can only be cleared when displaying either a Peak High or Peak Low recorded pressure. Press the clear button for at least ½ second. Dashed lines will briefly appear across the display indicating that both Peak values have been cleared. Both Peak High and Peak Low values will then display the cur- rent applied pressure.
  • Page 7 Functions  Tare ( -DD, Dual-Line Display XP2i Only ) Tare is a constant value subtracted from the true pressure. For instance, if you were mixing gases by partial pressure, you might want to fill a tank to 1760 psi with air, then add another 440 psi of helium.
  • Page 8 Functions  Automatic Shut-off The XP2i has a shutoff timer and will turn off automatically after 20 minutes of non-operation. Pressing any button or sending any command via the RS-232 connection resets the shutoff timer for another 20 minutes of operation. The XP2i will briefly display Auto Off 20 when turned on.
  • Page 9 Functions  Water Density (Inches of Water) The following applies only to models where inches of water is a selectable pressure unit. As shipped from the factory, the XP2i is set to display inches of water corresponding to the density of water at 4° C (39.2° F). You may require a different water density for your application, so the XP2i can be set to use the density of water at 20°...
  • Page 10 Functions  Calibration Due Reminder The XP2i has a calibration reminder feature that can help to assure that you never use the gauge after its calibration certificate has expired. It can be pro- grammed to alert you prior to, on, and after the gauge’s calibration due date. Due dates, reminder times, and message types can be set through our free ConfigXP software.
  • Page 11 Functions  Many other battery types and models have been tested but failed to meet the requirements for Intrinsic Safety—do not assume other models are equivalent. The XP2i can be operated from an external power supply (AC adapter kit PN: 2984). WARNING: Do not use AC adapter in hazardous locations.
  • Page 12: Data Logging With Dataloggerxp

    Data Logging with DataLoggerXP  Data Logging with DataLoggerXP „ „ „ „ „ WHAT IS DATALOGGERXP? DataLoggerXP is an optional data logging mode available for the XP2i gauge. You may purchase the DataLoggerXP option with your original order or add it later. With the XP2i’s long battery life, and Ultra Low Power (ULP) mode, you can log data over an entire year, without the need for external power supplies or battery replacements.
  • Page 13 Data Logging with DataLoggerXP  DataLoggerXP Terminology Clear..... Removes all stored data from the XP2i memory. Individual runs cannot be removed. Download .
  • Page 14: Installing The Software

    Data Logging with DataLoggerXP  „ „ „ „ „ INSTALLING THE SOFTWARE DataLoggerXP can either be installed from a CD provided by Crystal Engineering, or downloaded from the Crystal Engineering website. We recommend you check the website for the latest version. Installing from the CD Run the Setup.exe application located on the install CD.
  • Page 15: Dataloggerxp Software Operation

    Data Logging with DataLoggerXP  „ „ „ „ „ DATALOGGERXP SOFTWARE OPERATION As described in the prior section, the DataLoggerXP application provides complete control of the settings for data logging mode on the XP2i. You can change logging parameters and download data using the application. In this section, you will find a description of each of the functional areas of the application. The DataLoggerXP Main Window Working in the Main Window Setting the Options...
  • Page 16 Data Logging with DataLoggerXP  To Connect to the Gauge Plug the serial cable into the computer and XP2i gauge ports. In the DataLoggerXP main window, click Connect. The program will retrieve the current logging parameters, and then allow downloading, viewing, and saving data, as well as updating the logging parameters.
  • Page 17 Data Logging with DataLoggerXP  Data Logging Parameters Logging parameters can be set up one of two ways: on a computer connected to an XP2i, or from the XP2i keypad. This second option allows the user to make changes in the logging parameters while in the field. Once the logging parameters have been selected, the gauge can begin recording data. The two param- eters that affect data logging (logging interval and logging type) are covered below.
  • Page 18 Data Logging with DataLoggerXP  Data Management To Download Data from the Gauge Connect the gauge to your computer. In the DataLoggerXP main window, click the Download button. Depending on the number of readings, it can take up to 15 seconds to download all the data. Once the data has been downloaded, the View and Save buttons will be available.
  • Page 19 Data Logging with DataLoggerXP  To Save a Data Run In either the DataLoggerXP main window or the View Data dialog box, click the Save button. ƒ Save button in the DataLoggerXP main window. Save button in the View Data dialog box. You may download one recording run at a time.
  • Page 20: Xp2I Dataloggerxp Gauge Operation

    Data Logging with DataLoggerXP  „ „ „ „ „ XP2I DATALOGGERXP GAUGE OPERATION When in data logging mode, XP2i provides data logging capability along with many of the features of a standard XP2i. In order to allow operation as a data logger, some of the standard functionality of the standard XP2i is suspended.
  • Page 21 Data Logging with DataLoggerXP  Keypad Usage The following buttons and button combinations are used to control data logging runs on the XP2i. Where two keys are indicated, the operation is executed by pressing and holding down the first key, then pressing and holding down the second key until the action is completed. Note: Start by turning off the gauge.
  • Page 22 Data Logging with DataLoggerXP  Units units Pressing and holding the units button briefly causes the XP2i to display information about the logging state. If you continue to hold the units button, a logging run will begin or end. Immediately after pressing the units button, the gauge will display the operating mode as one of these values: The logging memory is Clear, data is empty, and the XP2i is idle.
  • Page 23 Data Logging with DataLoggerXP  Ultra Low Power (ULP) Mode The XP2i can log for extended periods with Ultra Low Power Mode, activated by the key combinations explained in Keypad Usage on page 19. Ultra Low Power Mode takes readings in Actual Mode, according to the set logging interval. The letters replace the live pressure reading during a recording.
  • Page 24: Excel Templates

    Data Logging with DataLoggerXP  „ „ „ „ „ EXCEL TEMPLATES In order to provide for a flexible reporting solution, DataLoggerXP is capable of directly interfacing with Microsoft® Excel to paste run data for reporting and graphing. DataLoggerXP supports Excel 2003, 2007, and 2010. This section assumes some familiarity with Excel. Note: For information on how to save data to the different file types see To Save a File in Excel Format on page 17.
  • Page 25: Enclosure

    Enclosure  Enclosure (U.S. Patent D612277) „ „ „ „ „ INFORMATION Weight ......562 g (19.8 oz) (includes batteries) Housing .
  • Page 26: Serial Numbers

    Enclosure  „ „ „ „ „ SERIAL NUMBERS Serial Number Location The serial number of your XP2i is located in two places; on the stem above the pressure fitting and behind the battery cover under the battery furthest from the serial port connection.
  • Page 27: Specifications

    Specifications  Specifications Accuracy – Gauge Pressure Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. Gauges must be exercised whenever exposed to significant changes in environmental conditions to achieve these specifications, and then re-zeroed. To exercise a gauge, cycle the gauge between zero (ambient barometric pressure) and the pressure of interest.
  • Page 28 Specifications  Accuracy – Absolute Pressure Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. Gauges must be exercised whenever exposed to significant changes in environmental conditions to achieve these specifications. To exercise a gauge, cycle the gauge between zero (ambient barometric pressure) and the pressure of interest.
  • Page 29 Specifications  Media Compatibility Liquids and gases compatible with 316 Stainless Steel. Pressure Conversions 1 psi = 27.6806 inches of water column (water at 4° C [39.2° F]) 0.070307 kilograms per square centimeter 27.7070 inches of water column (water at 15.6° C [60° F]) 68.948 millibar 27.7292 inches of water column (water at 20°...
  • Page 30: Power

    Specifications  „ „ „ „ „ POWER Batteries ......Three size AA (LR6) batteries. WARNING: Do not remove or change the batteries in hazardous locations. Approved Batteries ATEX/IECEx: The XP2i is Intrinsically Safe only if powered by one of the following battery types:...
  • Page 31: Pressure Ranges, Display Scales, & Resolution

    Specifications  „ „ „ „ „ PRESSURE RANGES, DISPLAY SCALES, & RESOLUTION in H Overpressure inHg mmHg kg/cm mbar 15PSI 1BAR  100KPA  6.5 x 0.001 0.01 0.001 0.01 0.0001 0.0001 0.01 0.00001 30PSI  2BAR 200KPA  3.0 x 0.001 0.01 0.001 0.0001 0.0001...
  • Page 32: Safety & Certifications

    Safety & Certifications  Safety & Certifications „ „ „ „ „ HAZARDOUS LOCATIONS Every XP2i pressure gauge includes the following Intrinsic Safety approvals: II 1G Ex ia IIC T4/T3 Ga Exia Intrinsically Safe and Non-Incendive for Hazardous Locations: FTZU 12 ATEX 0048 X Class I, Division 1, Groups A, B, C and D, Temperature Code T4/T3C.
  • Page 33: Atex/Iecex Safety Instructions

    Safety & Certifications  „ „ „ „ „ ATEX/IECEX SAFETY INSTRUCTIONS Bezpečnostní Instrukce Pro Prostředí s Nebezpečím Výbuchu – ČESKY (Czech) • V prostředí s nebezpečím výbuchu nepoužívejte přípojku RS-232. • Baterie vyměňujte pouze v bezpečném prostředí. Používejte pouze schválené baterie. •...
  • Page 34 Safety & Certifications  Safety Instructions for Hazardous Locations – ENGLISH (English) • Do not use the RS-232 connector in a hazardous location. • Replace batteries in non-hazardous locations, with approved batteries, only. • It is the users responsibility to understand the proper application of this product in potentially explosive atmospheres. •...
  • Page 35 Safety & Certifications  Instructions De Sécurité Pour Les Zones Dangereuses – FRANÇAIS (French) • Ne pas utilisez le connecteur RS-232 dans une Zone Dangereuse. • Remplacez les piles dans des Zones non-dangereuses, avec les piles appropriées, uniquement. • Il est de la responsabilité de l’utilisateur de bien comprendre l’application appropriée de ce produit en atmosphères explosives. •...
  • Page 36 Safety & Certifications  Prescrizioni Di Sicurezza Per Area Pericolosa – ITALIANO (Italian) • Non utilizzare il connettore Rs-232 in Area Pericolosa. • Sostituire le batterie in Aree non Pericolose e solamente con Batterie approvate. • E’ responsabilità dell’utilizzatore comprendere l’adatta applicazione di questo prodotto in atmosfere potenzialmente esplosive. •...
  • Page 37 Safety & Certifications  Instrukcja Bezpieczeństwa Dla Srefy Zagrożonej Wybuchem– POLSKI (Polish) • Połączenie RS232 może być używane tylko poza strefą zagrożenia wybuchem. • Wymiana baterii tylko poza strefą zagrożenia wybuchem, używaż tylko zatwierdzony typ baterii. • Odpowiedzialnością użytkownika jest używanie tego produktu we wlaściwy sposób w strefie zagrożonej wybuchem. •...
  • Page 38: Support

    Support  Support „ „ „ „ „ TROUBLESHOOTING The XP2i is a very high performance gauge. Due to the high resolution of this product, you may observe conditions that appear to be defects in the product, but are in fact a result of being able to resolve and measure pressure to a degree not possible with other instruments. Noisy or Unstable Reading When Used with Fluids Err 2 Displayed When calibrating or comparing the indicated pressure from an XP2i against a hydraulic dead...
  • Page 39: Calibration

    Support  „ „ „ „ „ CALIBRATION Factory Adjustment If adjustment is required, we recommend returning the unit to the factory. Factory service offers benefits you won’t find anywhere else. We have the facili- ties to test your gauge at a variety of temperatures utilizing NIST traceable standards in our A2LA accredited laboratory (#2601.01), resulting in calibration certificates that provide performance data over temperature.
  • Page 40: Software

    Support  „ „ „ „ „ SOFTWARE Programming Instructions Programming Instructions. Control and communicate with XP2is using our Integrate XP2is into your test environment! ConfigXP Configuration Software ConfigXP to disable unwanted pressure units, set default pressure units, change water density, adjust calibration, and more via the RS-232 interface. „...
  • Page 41: Contact Us

    China Crystal Engineering • California Tel +44 (0)1243 833 302 Tel +49 (0)2159 9136 510 Tel +91 22 2836 4750 Shanghai jofra@ametek.co.uk info.mct-de@ametek.de jofra@ametek.com Tel +1 (800) 444 1850 Tel +86 21 5868 5111 Fax +1 (805) 595 5466 Beijing...
  • Page 42 © 2018 Crystal Engineering Corporation 708 Fiero Lane, Suite 9, San Luis Obispo, California 93401-8701 4870. Q...

Table of Contents