  • Page 1 HPC50 Series Operation Manual for the HPC50 Series Intrinsically Safe Handheld Pressure Calibrator...
  • Page 2: Table Of Contents

    Contents Overview Enclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
  • Page 3: Overview

    INTRODUCTION Thank you for choosing the HPC50 Series intrinsically safe handheld pressure calibrator from Crystal Engineering Corporation. HPC50 calibrators feature dead- weight tester accuracy in a modern digital package with the safest fitting connection available. The full color display features an easy to use interface and may be used in a broad range of applications from simple tool type jobs to complex calibration jobs in custody transfer systems.
  • Page 4: Quickstart

    All input and output Easy to use single layer user interface. No deep menu structure! connectors are placed Operate and set up the HPC50 Series to perform your tasks quickly away from the display and and intuitively. keypad to give maximum freedom to operate.
  • Page 5: Functions

    Press and hold the power button to turn the HPC50 Series on or off. The HPC50 will automatically power down if not used for the time period defined in the unit settings or in CrystalControl. Automatic Shutoff - Low Power Mode To optimize battery life, adjust your Automatic Shutoff time (shut off time in absence of key press) on the unit or through the free CrystalControl software.
  • Page 6 Note: Configurable fields are not highlighted in light blue on the actual display; but selected fields are highlighted in dark blue. Use the enter button to select a field for editing. HPC50 Series Operation Manual...
  • Page 7 Calibration Due Reminder The HPC50 Series has a calibration reminder feature that can help to assure that you never use the calibrator after its calibration certificate has expired. Set calibration due dates and warning period through our free CrystalControl software.
  • Page 8: Settings

    Maximum Resolution: When a resolution less than standard is selected, the displayed number will include an asterisk ( • Overpressure Count: Counts the times when pressure has exceeded 110% of the range • View Calibration Details • Serial Number • Part Number HPC50 Series Operation Manual...
  • Page 9 When finished press the enter button. The name is limited to eight characters. Highlight the settings to modify... then change the name of the saved setup. Press the function button for Save. HPC50 Series Operation Manual...
  • Page 10: Display

    The display is divided into two sections: the upper and lower displays. The upper display can be configured to show values for Current, Voltage, Temperature, Pressure, BARO, or Switch Test. The lower display can be configured to show values for only Temperature, Pressure, and BARO. Upper Display (showing Current) Lower Display (showing Pressure) HPC50 Series Operation Manual...
  • Page 11: Operation

    PRESSURE SENSORS The HPC50 Series is equipped with one or two pressure sensors. In addition, one or two intrinsically safe APMi (Advanced Pressure Module) or ATMi (Advanced Temperature Module) external modules may also be used. For information on the Crystal external modules, please refer to APMi Modules ATMi Modules.
  • Page 12  Overpressure Conditions The HPC50 Series will read pressure up to approximately 110% of the rated pressure range. Above 110%, the display will indicate OL, and readings will stop updating. The zero function does not affect when the display will indicate...
  • Page 13 Never disconnect pressure instrumentation without first relieving system pressure! Differential Pressure The HPC50 Series can display differential pressure when 2 or more sensors are installed (including external APMi). The modules do not have to be the same full scale pressure range.
  • Page 14 ( ) icon for four seconds until the differential display readings change from (- - - - -) to (- - -). The tare indication will now be removed from the display. HPC50 Series Operation Manual...
  • Page 15: Measuring Current

    The HPC50 will display the measured current. Note: The display will indicate + OL if the measured current exceeds the nominal range of current measurement (55 mA). Note: For details on % error, % flow, and scaling features see Percent Error Function. HPC50 Series Operation Manual...
  • Page 16: Sourcing Current

    Or, if you would like to pause the step function and increase manually, press Stop. If you press Stop, new function keys -Step (decreasing a single step) and +Step (increasing a single step) are now available. HPC50 Series Operation Manual...
  • Page 17: Measuring Voltage

    For more information on working in the HPC50 display, see Navigating Through a Typical Setup. Connect the HPC50 using the appropriate test leads. The HPC50 will indicate the measured voltage. Note: + OL The display will indicate if the measured voltage exceeds 30 VDC. HPC50 Series Operation Manual...
  • Page 18: Measuring Temperature (Atmi)

    For details on the ATMi external module, see the ATMi External Temperature Modules CAUTION: ATMi modules must only be used with intrinsically safe probes provided by Ametek. To Measure Temperature Connect an ATMi with RTD probe to the HPC50 Chassis in either ATMi port.
  • Page 19: Performing A Switch Test

    Dead band values will be indicated. Switch Open Switch Closed To perform a new test, highlight the reset option and press the enter button. The Closed, Opened, and Dead band values will all be blank again. HPC50 Series Operation Manual...
  • Page 20: Calibrating A Pressure Transmitter

    0 and 100% points for both the mA and pressure ranges of the transmitter. Increase pressure to the desired point and the HPC50 upper display will indicate the % error as well as the actual mA value. The lower display will show the measured pressure. HPC50 Series Operation Manual...
  • Page 21 Increase pressure to the desired point and the HPC50 upper display will indicate the equivalent pressure based on the mA reading, as well as the actual mA value. The lower display will show the measured pressure. HPC50 Series Operation Manual...
  • Page 22: Enclosure

    SENSOR DIAPHRAGM SURFACE 1/4 INCH MNPT 1/4 INCH MBSP M20X1.5 ADAPTER ADAPTER ADAPTER 22.7 (0.90) 7/16-20 MP CPF PORTS 26.1 36.5 19.1 29.1 32.6 (1.03) (1.44) (0.75) (1.15) (1.30) ADDITIONAL SENSOR LENGTH WITH OPTIONAL FITTING ADAPTERS HPC50 Series Operation Manual...
  • Page 23: Apmi Modules

    30 to 110% of Range: ... . ±(0.035% of Reading) Temperature Range ..-20 to 50° C (-4 to 122° F) The BARO option for the HPC50 Series allows you to toggle between Vacuum : .
  • Page 24: Atmi Modules

    6 PIN CONNECTION Combine with part number 127387 for a -45 to 150° C temperature specification . FOR HPC50 SERIES CABLE sensor . Contact us to add a calibration certificate . Applies to all modules . „ „ „ „ „...
  • Page 25: Specifications

    Exposure to environmental extremes of temperature, shock, or vibration may warrant a more frequent calibration period. The HPC50 Series must be exercised and re-zeroed whenever exposed to significant changes in environmental conditions to achieve these specifications. To exercise a gauge, cycle the gauge between zero (ambient barometric pressure) and the pressure of interest. A properly exercised gauge will return to a zero reading (or return to the same ambient barometric reading).
  • Page 26 Diaphragm Seal Fluid: ... . Silicone Oil Connection: ..... Crystal CPF Female HPC50 Series Operation Manual...
  • Page 27 Voltage Burden @ 50mA ..< 0.250 V HART Resistor ....250 Ω HPC50 Series Operation Manual...
  • Page 28 Display Rate ..... 3 readings/second (standard) Operating and Storage Temperature Operating Temperature Range . . . -20 to 50° C (-4 to 122° F) Storage Temperature Range ..-40 to 75° C (-40 to 167° F) HPC50 Series Operation Manual...
  • Page 29 The HPC50 uses three AA batteries. Unscrew the four flathead screws to gain access to the battery compartment. Replace the batteries taking care to note po- larity for their proper installation. Failure to properly seal the battery compartment may allow water damage that could permanently compromise the HPC50. HPC50 Series Operation Manual...
  • Page 30: Ranges And Resolutions

    1/4 NPT ..(omit) and BARO option. HPC52 ..(Dual) Yes . . . - BARO G 1/4 B ..-BSP M20x1.5 . . . -M20 HPC50 Series Operation Manual...
  • Page 31: Safety & Certifications

    „ „ „ „ „ CERTIFICATIONS The HPC50 Series has been tested and certified to comply with a variety of international standards. This HPC50 complies with the Australian Radiocommunications (Electromagnetic Compatibility) Standard 2008. Crystal Engineering declares that the HPC50 is in accordance with the ATEX Directive, the Electromagnetic Compatibility Directive, and the Pressure Equipment Directive per our declaration(s).
  • Page 32: Support

    TROUBLESHOOTING The HPC50 Series is a very high performance calibrator. Due to the high resolution of this product, you may observe conditions that appear to be defects in the product, but are in fact a result of being able to read and measure pressure to a degree not possible with other instruments.
  • Page 33: Calibration

    CAUTION: Do not subject the BARO sensor to pressures less than 700 mbarA (10.153 PSIA), or greater than 1100 mbarA (15.954 PSIA), as this may cause permanent damage. Use only clean a clean dry pneumatic source. CAUTION: Direct contact with the surface of the BARO sensor may cause permanent damage. Direct sunlight on exposed BARO sensor may affect readings slightly. HPC50 Series Operation Manual...
  • Page 34: Accessories And Replacement Parts

    Hard plastic carrying case with molded eggshell foam interior . MPM-1/4MPT CPF Male to 1/4" Male NPT Fitting (Included as standard) MPM-1/4BSPM CPF Male to 1/4” Male BSP Fitting (Included with -BSP) MPM-M20x1.5M CPF Male to M20 Male Adapter (Included with -M20) HPC50 Series Operation Manual...
  • Page 35: Contact Us

    WARRANTY Crystal Engineering Corporation warrants the HPC50 Series Calibrator to be free from defects in material and workmanship under normal use and service for one (1) year from date of purchase to the original purchaser. It does not apply to batteries or when the product has been misused, altered or damaged by ac- cident or abnormal conditions of operation.
  • Page 36 © 2019 Crystal Engineering Corporation 708 Fiero Lane, Suite 9, San Luis Obispo, California 93401-8701 6486. A 1903...