Specifications
Gauge Pressure Accuracy (P1, P2, and APMi)
Includes all effects of linearity, hysteresis, repeatability, temperature, and stability with the operating temperature range for one year.
Note:
Exposure to environmental extremes of temperature, shock, or vibration may warrant a more frequent calibration period.
The HPC50 Series must be exercised and re-zeroed whenever exposed to significant changes in environmental conditions to achieve these specifications. To
exercise a gauge, cycle the gauge between zero (ambient barometric pressure) and the pressure of interest. A properly exercised gauge will return to a zero
reading (or return to the same ambient barometric reading).
Gauge Pressure (psi)
X
•
18 to 28° C
0 to 30% of Range . . . . . . . . . . . . . . . . . . ±(0.01% of Full Scale)
30 to 110% of Range . . . . . . . . . . . . . . . . ±(0.035% of Reading)
*
Vacuum
. . . . . . . . . . . . . . . . . . . . . . . . . . . ±(0.05% of Full Scale
•
-20 to 50° C
0 to 30% of Range . . . . . . . . . . . . . . . . . . ±(0.015% of Full Scale)
30 to 110% of Range . . . . . . . . . . . . . . . . ±(0.050% of Reading)
*
Vacuum
. . . . . . . . . . . . . . . . . . . . . . . . . . . ±(0.05% of Full Scale
*
Applies to 300 psi and lower ranges only.
Vacuum Range = -14.5 psi.
**
Full Scale is the numerical value of the positive pressure range.
!
CAUTION: Not recommended for continuous use below -14.5 psig. Refer to the
vacuum use.
Absolute Pressure with BARO Option (psiA)
X
All absolute accuracies are equivalent to the gauge pressure accuracies, except as noted below.
15 psi Range . . . . . . . . . . . . . . . . . . . . . . . Gauge Accuracy + 0.005 psiA
30 psi Range . . . . . . . . . . . . . . . . . . . . . . . Gauge Accuracy + 0.005 psiA
100 psi Range . . . . . . . . . . . . . . . . . . . . . . Gauge Accuracy + 0.002 psiA
**
)
**
)
XP2i-DP
data sheet for gauges that are intended for continuous high
Specifications
23
HPC50 Series Operation Manual