Ametek Crystal HPC51 Operation Manual page 29

Intrinsically safe handheld pressure calibrator
Table of Contents

Advertisement

Power
The HPC50 can be powered by batteries or USB.
!
WARNING: Do not use the USB port in a hazardous area.
Power Icon States
X
The HPC50 battery icon indicates how much battery life is remaining.
100%
50%
0%
Battery Power
X
The HPC50 uses three(3) size AA (LR6) alkaline batteries.
Note:
If the batteries discharge too deeply, the calibrator will automatically shut down to avoid battery leakage and false measurements.
Approved Batteries
CSA
Approved
Temperature
Battery Type
Class
Rayovac Max Plus 815
T3
Energizer
E91*
T4
Duracell MN1500
T3
Energizer
EN91*
Replace batteries with approved type in non-hazardous locations only.
*
Energizer is manufactured by Energizer Holdings, Inc., and the Eveready Battery Company, Inc.
Battery Life
X
Depending on how the HPC50 is configured, the battery life may differ. All numbers in the table below are without the use of the backlight, which will de-
crease the battery life.
Configuration
Battery Life, In Hours
HPC50 with no external modules
HPC50 with one external module
HPC50 with two external modules
Battery Replacement
X
The HPC50 uses three AA batteries. Unscrew the four flathead screws to gain access to the battery compartment. Replace the batteries taking care to note po-
larity for their proper installation. Failure to properly seal the battery compartment may allow water damage that could permanently compromise the HPC50.
ATEX & IECEx
Ambient
Temperature
Ambient
Temperature
Class
Temperature
-20 to 50° C
T4
-20 to 50° C
-20 to 40° C
-20 to 40° C
T3
-20 to 50° C
35
25
12
27
Specifications
HPC50 Series Operation Manual