Atmi Modules - Ametek Crystal HPC51 Operation Manual

Intrinsically safe handheld pressure calibrator
Table of Contents

Advertisement

ATMi Modules

       
TEMPERATURE MEASUREMENT
Accuracy: . . . . . . . . . . . . . . . . . . . . . . . . ±(0.015% of rdg) + 0.02 Ohm
Range: . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 400 Ohms
Resolution: . . . . . . . . . . . . . . . . . . . . . . 0.01 on all scales
Units: . . . . . . . . . . . . . . . . . . . . . . . . . . . . °C, K, °F, R, Ω
TCR: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.003850 Ω/Ω/°C (IEC 60751)
Wiring: . . . . . . . . . . . . . . . . . . . . . . . . . . 4-wire support
Includes all effects of linearity, hysteresis, repeatability, temperature,
and stability for one year .
Combine with part number 127387 for a -45 to 150° C temperature
sensor . Contact us to add a calibration certificate .
       
OUTPUT
Temperature Resolution . . . . . . . . . 0.01
Display Update . . . . . . . . . . . . . . . . . . up to 10 per second
Temperature resolution and display update are the maximum values
available . The resolution of your Crystal device may be different .
The proper selection of the RTD sensing element is very important as the error associated with this device is the majority of the overall system measurement uncertainty . IEC 751 is the
standard that defines the temperature versus resistance for 100Ω, 0 . 0 0385 Ω/Ω/°C platinum RTDs . IEC 751 defines two classes of RTDs: Class A and B . Class A RTDs operate over the -200 to
630°C range versus -200 to 800°C for the Class B elements . For example, the Class A uncertainty is about half that of the Class B elements as illustrated in the following table .
HPC50/ATMi
Class A
Temperature
Uncertainty
Uncertainty
°C
±Ω
±°C
±Ω
-200
0.02
0.05
0.24
-40
0.03
0.08
0.09
0
0.04
0.09
0.06
50
0.04
0.10
0.10
100
0.04
0.11
0.13
150
0.04
0.12
0.17
200
0.05
0.13
0.20
400
0.06
0.17
0.33
600
0.07
0.21
0.43
800
0.08
0.25
0.52
       
COMMUNICATION
Connector . . . . . . . . . . . . . . . . . . . . . . . 6-pin LEMO
Serial . . . . . . . . . . . . . . . . . . . . . . . . . . . . RS-422, 9600 baud, 8 data,
Protocol . . . . . . . . . . . . . . . . . . . . . . . . . ASCII command language
       
OPERATING TEMPERATURE
Temperature Range . . . . . . . . . . . . . . -20 to 50° C (-4 to 122° F)
< 95% RH, non-condensing . No change in accuracy over operating
temperature range . Gauge must be zeroed to achieve rated
specification .
Applies to all modules .
       
STORAGE TEMPERATURE
Temperature Range . . . . . . . . . . . . . . -40 to 75° C (-40 to 167° F)
Class A
HPC50/ATMi + Class A
Class B
Uncertainty
Uncertainty
±°C
±Ω
±°C
±Ω
0.55
0.24
0.55
0.56
0.23
0.10
0.24
0.20
0.15
0.07
0.17
0.12
0.25
0.10
0.27
0.21
0.35
0.14
0.37
0.30
0.45
0.17
0.46
0.39
0.55
0.21
0.56
0.48
0.95
0.33
0.96
0.79
1.35
0.44
1.37
1.06
1.75
0.53
1.77
1.28
       
ENCLOSURE
Dimensions . . . . . . . . . . . . . . . . . . . . . . 2.5 x 1.1 in (63.3 x 27.0 mm)
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . 0.31 lbs (141.0 g)
no parity, 1 stop
6 PIN CONNECTION
Class B
HPC50/ATMi + Class B
Uncertainty
±°C
±Ω
±°C
1.30
0.56
1.30
0.50
0.20
0.51
0.30
0.12
0.31
0.55
0.22
0.56
0.80
0.31
0.81
1.05
0.39
1.06
1.30
0.48
1.31
2.30
0.79
2.31
3.30
1.06
3.31
4.30
1.28
4.31
FOR REMOTE
TEMPERATURE
PROBE
63.3
(2.49)
6 PIN CONNECTION
FOR HPC50 SERIES CABLE
HPC50 Series Operation Manual
ATMi Modules
22
Ø 27
(1.06)

Advertisement

Table of Contents
loading

Table of Contents